Knowledge (XXG)

Grid dip oscillator

Source 📝

195: 36: 106: 244:
The operator adjusts the frequency of the GDO until its meter shows its lowest reading (the "dip"). The frequency is read from the dial on the GDO, or the frequency can be measured by finding the dip meter's signal on a well-calibrated radio receiver. Some modern GDOs have a built-in frequency meter,
231:
The coil and the test circuit can either be inductively or capacitively coupled: Coupling is inductive if the coil wires are held parallel to the nearest wires of the circuit being tested, capacitive if the coil wires and the circuit wires are held perpendicular. Depending on the context desired for
210:
a nearby circuit (in which case the amplitude shown on the meter "dips") or to measure the relative power absorbed from a nearby powered circuit (in which case, the meter amplitude peaks). In either mode of operation, some experimentation is needed to find a distance between the pickup coil and the
85:
that circulates a small-amplitude signal through an exposed coil, whose electromagnetic field can interact with adjacent circuitry. The oscillator loses power when its coil is near a circuit that resonates at the same frequency. A meter on the GDO registers the amplitude drop, or "dip", hence the
223:
whose exposed external coil will loosely couple to the measured resonant circuit when held moderately close by. The measuring distance must be skillfully adjusted to be close enough to provide sufficient coupling for the dip to show clearly, but far enough that the meter and the tested circuit
240:
In ordinary use, only the oscillator in the dip meter is powered, and the circuit under test's only power is what it drains from the signal in the GDO coil. When both circuits are resonant at the same frequency, power transfer from the coil to the adjacent tested circuit reaches a maximum,
261:
in meter power occurs in the unpowered GDO, when its coil is held near the wires of an active resonant circuit. Since the power in the circuit under test must be high enough to register on the meter, this uncommon method is risky both for the operator and the equipment.
131:
whose coupling is most effective when their respective resonant frequencies match. The degree of coupling affects the frequency and amplitude of oscillation in the dip meter, which is sensed in any of several ways, the simplest and most usual of which is a built-in
211:
circuit under test, to ensure that the two circuits are close enough to transfer power, but not coupled so closely that the circuit supplied with power overwhelms the responding circuit, and forces it to oscillate regardless of the frequency.
232:
the measurement, the circuit under test can be temporarily disconnected from its surroundings to avoid distortion by the parts it is normally attached to, or left wired in place to measure the response of the combined system.
97:. They can also be used for transmission line testing, as signal generators, and for measuring inductance and capacitance of components. Measurement with a GDO is called "dipping" a circuit. 121:
and matching interchangeable coils, as shown in the circuit diagram. Resonance is indicated by a dip in amplitude of the signal within the GDO, by a meter on the device.
389: 224:
oscillate independently, so neither device's frequency is significantly distorted by the other, and so that power lost to the external circuit does not
175:
whose current is measured instead of a vacuum tube grid. Solid-state versions of the grid dip oscillator are more versatile, since they can operate at
341: 313: 443: 330: 448: 82: 241:
consequently the dip meter's oscillator amplitude reaches a minimum due to the power lost to the circuit under test.
453: 160: 124:
When the oscillator's exposed coil is in the vicinity of another resonant circuit, the coupled pair behaves as a
186:, which are more complex, but perform many of the same functions more conveniently, and require less dexterity. 136:. The distance between the coil and the tested circuit needs to be adjusted carefully so that the GDO amplitude 438: 383: 90: 458: 194: 271: 254: 118: 105: 39:
A dipmeter and several accessory probe coils. (Mitamusenkenkyūsho 三田無線研究所株式会社 DELICA DMC-230S2)
35: 371: 309: 183: 23: 94: 75: 67: 114: 432: 19: 153: 27: 225: 149: 128: 220: 172: 78: 406:. Komagane, Japan: Kenwood Electronic Inc. 24 January 2005. pp. 1, 7–10. 375: 71: 148:
Grid dip oscillators were first developed in the 1920s and were built with
30:
angle of Earth's magnetic field, the field line angle in a vertical plane.
199: 176: 140:
significantly affected by the coupled circuit, but its frequency is not.
125: 133: 366:
Hallas, Joel R. W1ZR (August 2016). "Antenna analyzers - the basics".
253:
Some dip meters can be used in reverse, as ultra-short-range, tuned
206:
The dip meter can be used either to measure the relative power lost
193: 104: 421: 93:
for measuring the properties of resonant circuits, filters, and
370:. Newington, CT: American Radio Relay League. pp. 32–34. 18:"Dip meter" can also refer to an influential early commercial 308:(90th ed.). Newington, CT: American Radio Relay League. 179:
and lower amplitude, and are not tethered by a power cord.
257:. The operator finds the frequency at which the highest 109:
Oscillator portion of a transistor-based gate-dip meter
329:
Houck, Harry W.; Gaw, Norman W. Jr. (December 1961).
245:
which makes over-coupling somewhat less troublesome.
152:. The devices displayed the amplitude of the tube's 26:; or may refer to an instrument that measures the 182:Dip meters of all types are being supplanted by 117:variable-frequency oscillator with a calibrated 306:The 2013 ARRL Handbook for Radio Communications 331:"New method of accurate frequency measurement" 8: 388:: CS1 maint: numeric names: authors list ( 299: 297: 295: 293: 291: 202:Grid Dip Meters with a set of tuning coils 171:in reference to the analogous part of the 219:The heart of the instrument is a tunable 89:Dip oscillators have been widely used by 338:Proceedings of the Radio Club of America 34: 287: 381: 404:Kenwood DM-81 Dip Meter Users' Manual 347:from the original on 18 February 2014 7: 163:devices, and are sometimes called 14: 113:Central to the dip meter is a 1: 304:Silver, H. Ward, ed. (2012). 83:variable-frequency oscillator 340:. Vol. 37, no. 4. 228:the dip meter's oscillator. 475: 249:Absorption frequency meter 66:, is a type of electronic 444:Electrical test equipment 236:Emission frequency meter 169:emitter dip oscillators 91:amateur radio operators 74:of nearby unconnected 449:Electronic oscillators 203: 159:Modern dip meters are 110: 101:Principle of operation 40: 255:field strength meters 197: 108: 38: 272:Absorption wavemeter 165:gate dip oscillators 156:current, hence GDO. 44:Grid dip oscillator 424:. picture gallery. 204: 111: 72:resonant frequency 70:that measures the 41: 454:Radio electronics 422:"Grid Dip Meters" 315:978-0-87259-419-7 215:Coupling distance 184:antenna analyzers 466: 425: 408: 407: 400: 394: 393: 387: 379: 363: 357: 356: 354: 352: 346: 335: 326: 320: 319: 301: 119:tuning capacitor 24:Dipmeter Advisor 474: 473: 469: 468: 467: 465: 464: 463: 429: 428: 420: 417: 412: 411: 402: 401: 397: 380: 365: 364: 360: 350: 348: 344: 333: 328: 327: 323: 316: 303: 302: 289: 284: 268: 251: 238: 217: 192: 146: 103: 76:radio frequency 50:), also called 12: 11: 5: 472: 470: 462: 461: 456: 451: 446: 441: 431: 430: 427: 426: 416: 415:External links 413: 410: 409: 395: 358: 321: 314: 286: 285: 283: 280: 279: 278: 275: 274: 267: 264: 250: 247: 237: 234: 216: 213: 191: 188: 145: 142: 115:high-frequency 102: 99: 79:tuned circuits 56:gate dip meter 52:grid dip meter 33: 32: 13: 10: 9: 6: 4: 3: 2: 471: 460: 457: 455: 452: 450: 447: 445: 442: 440: 439:Amateur radio 437: 436: 434: 423: 419: 418: 414: 405: 399: 396: 391: 385: 384:cite magazine 377: 373: 369: 362: 359: 343: 339: 332: 325: 322: 317: 311: 307: 300: 298: 296: 294: 292: 288: 281: 277: 276: 273: 270: 269: 265: 263: 260: 256: 248: 246: 242: 235: 233: 229: 227: 222: 214: 212: 209: 201: 196: 189: 187: 185: 180: 178: 174: 170: 166: 162: 157: 155: 151: 143: 141: 139: 135: 130: 127: 122: 120: 116: 107: 100: 98: 96: 92: 87: 84: 80: 77: 73: 69: 65: 61: 57: 53: 49: 45: 37: 31: 29: 25: 21: 20:expert system 16: 15: 459:Vacuum tubes 403: 398: 367: 361: 349:. Retrieved 337: 324: 305: 258: 252: 243: 239: 230: 218: 207: 205: 181: 168: 164: 158: 150:vacuum tubes 147: 137: 134:microammeter 123: 112: 88: 63: 59: 55: 51: 47: 43: 42: 28:magnetic dip 17: 351:12 November 161:solid-state 129:transformer 433:Categories 282:References 221:LC circuit 173:transistor 81:. It is a 68:instrument 62:, or just 376:0033-4812 190:Operation 60:dip meter 342:Archived 266:See also 200:Heathkit 177:higher Q 95:antennas 144:History 22:called 374:  312:  86:name. 64:dipper 345:(PDF) 334:(PDF) 226:swamp 126:low-Q 390:link 372:ISSN 353:2021 310:ISBN 259:rise 198:Two 154:grid 368:QST 167:or 48:GDO 435:: 386:}} 382:{{ 336:. 290:^ 208:to 138:is 58:, 54:, 392:) 378:. 355:. 318:. 46:(

Index

expert system
Dipmeter Advisor
magnetic dip

instrument
resonant frequency
radio frequency
tuned circuits
variable-frequency oscillator
amateur radio operators
antennas

high-frequency
tuning capacitor
low-Q
transformer
microammeter
vacuum tubes
grid
solid-state
transistor
higher Q
antenna analyzers

Heathkit
LC circuit
swamp
field strength meters
Absorption wavemeter

Text is available under the Creative Commons Attribution-ShareAlike License. Additional terms may apply.