Knowledge (XXG)

Isotope-ratio mass spectrometry

Source đź“ť

426:
with a higher acceleration potential (several 1000 V) in order to minimize the energy distribution of the ion beam. Modern instruments operate at 6-10kV. The radius of deflection of an ion within a magnetic field depends on the kinetic energy and the mass/charge ratio of the ion (strictly, the magnet is a momentum analyzer not just a mass analyzer). Because of the large energy distribution, ions with similar mass/charge ratio can have very different kinetic energies and will thus experience different deflection for the same magnetic field. In practical terms one would see that ions with the same mass/charge ratio focus at different points in space. However, in a mass-spectrometer one wants ions with the same mass/charge ratio to focus at the same point, e.g. where the detector is located. In order to overcome these limitations, commercial MC-ICP-MS are double-focusing instruments. In a double-focusing mass-spectrometer ions are focused due to kinetic energy by the ESA (electro-static-analyzer) and kinetic energy + mass/charge (momentum) in the magnetic field. Magnet and ESA are carefully chosen to match the energy focusing properties of one another and are arranged so that the direction of energy focusing is in opposite directions. To simplify, two components have an energy focus term, when arranged properly, the energy term cancels out and ions with the same mass/charge ratio focus at the same point in space. It is important to note, double-focusing does not reduce the kinetic energy distribution and different kinetic energies are not filtered or homogenized. Double-focusing works for single as well as multi-collector instruments. In single collector instruments ESA and magnet can be arranged in either forward geometry (first ESA then magnet) or reversed geometry (magnet first then ESA), as only point-to-point focusing is required. In multi-collector instruments, only forward geometry (ESA then magnet) is possible due to the array of detectors and the requirements of a focal plane rather than a focal point.
255:) by means of a system of valves, so that a number of comparison measurements are made of both gases. In continuous flow IRMS, sample preparation occurs immediately before introduction to the IRMS, and the purified gas produced from the sample is measured just once. The standard gas may be measured before and after the sample or after a series of sample measurements. While continuous-flow IRMS instruments can achieve higher sample throughput and are more convenient to use than dual inlet instruments, the yielded data is of approximately 10-fold lower precision. 185:(heavy hydrogen) is almost double the mass of the common hydrogen isotope. Water molecules containing the common hydrogen isotope (and the common oxygen isotope, mass 16) have a mass of 18. Water incorporating a deuterium atom has a mass of 19, over 5% heavier. The energy to vaporise the heavy water molecule is higher than that to vaporize the normal water so isotope fractionation occurs during the process of evaporation. Thus a sample of sea water will exhibit a quite detectable isotopic-ratio difference when compared to Antarctic snowfall. 38: 442: 114: 353: 1418: 216: 1442: 1430: 504:(AMS) can be used. For example, the decay rate of the radioisotope C is widely used to date organic materials, but this approach was once limited to relatively large samples no more than a few thousand years old. AMS extended the range of C dating to about 60,000 years BP, and is about 10 times more sensitive than conventional IRMS. 379:
SIMS is a common method used in U-Pb analysis, as the primary ion beam is used to bombard the surface of a single zircon grain in order to yield a secondary beam of Pb ions. The Pb ions are analyzed using a double focusing mass spectrometer that comprises both an electrostatic and magnetic analyzer.
425:
The Ar-ICP produces an ion-beam with a large inherent kinetic energy distribution, which makes the design of the mass-spectrometer somewhat more complex than it is the case for conventional TIMS instruments. First, different from Quadrupole ICP-MS systems, magnetic sector instruments have to operate
507:
AMS works by accelerating negative ions through a large (mega-volt) potential, followed by charge exchange and acceleration back to ground. During charge exchange, interfering species can be effectively removed. In addition, the high energy of the beam allows the use of energy-loss detectors, that
263:
A static gas mass spectrometer is one in which a gaseous sample for analysis is fed into the source of the instrument and then left in the source without further supply or pumping throughout the analysis. This method can be used for 'stable isotope' analysis of light gases (as above), but it is
145:
design that this type of instrument is often called the 'Nier type'. In the most general terms the instrument operates by ionizing the sample of interest, accelerating it over a potential in the kilo-volt range, and separating the resulting stream of ions according to their mass-to-charge ratio
528:
and water by combustion. The gas stream finally enters a capillary, is dried, ionized, and analyzed. This process allows a mixture of compounds to be purified and analyzed continuously, which can decrease the analysis time by a factor of four. Moving wire IRMS is quite sensitive, and samples
392:) can be used. A SHRIMP is a double-focusing mass spectrometer that allows for a large spatial separation between different ion masses based on its relatively large size. For U-Pb analysis, the SHRIMP allows for the separation of Pb from other interfering molecular ions, such as HfO 153:
Many radiogenic isotope measurements are made by ionization of a solid source, whereas stable isotope measurements of light elements (e.g. H, C, O) are usually made in an instrument with a gas source. In a "multicollector" instrument, the ion collector typically has an array of
817:
Williams, I.S. (1998), "U-Th-Pb geochronology by ion microprobe", In: McKibben, M.A.; Shanks III, W.C.; Ridley, W.I.; (Editors), "Applications of microanalytical techniques to understanding mineralizing processes", Reviews in Economic Geology Special Publication 7:
319:
When these isotope ratios are measured by TIMS, mass-dependent fractionation occurs as species are emitted by the hot filament. Fractionation occurs due to the excitation of the sample and therefore must be corrected for accurate measurement of the isotope ratio.
323:
There are several advantages of the TIMS method. It has a simple design, is less expensive than other mass spectrometers, and produces stable ion emissions. It requires a stable power supply, and is suitable for species with a low ionization potential, such as
133:
type in this field of research for two reasons. First, it can be set up for multiple-collector analysis, and second, it gives high-quality 'peak shapes'. Both of these considerations are important for isotope-ratio analysis at very high precision and accuracy.
335:
The disadvantages of this method stem from the maximum temperature achieved in thermal ionization. The hot filament reaches a temperature of less than 2500°C, leading to the inability to create atomic ions of species with a high ionization potential, such as
421:
has normally been determined by TIMS. However, some systems (e.g. Hf-W and Lu-Hf) are difficult or impossible to analyse by TIMS, due to the high ionization potential of the elements involved. Therefore, these methods can now be analysed using MC-ICP-MS.
376:(SIMS). This type of ion-microprobe analysis normally works by focusing a primary (oxygen) ion beam on a sample in order to generate a series of secondary positive ions that can be focused and measured based on their mass/charge ratios. 227:
It is critical that the sample be processed before entering the mass spectrometer so that only a single chemical species enters at a given time. Generally, samples are combusted or pyrolyzed and the desired gas species (usually hydrogen
409:
analysis uses a quadrupole analyser, which only allows single-collector analysis. Due to the inherent instability of the plasma, this limits the precision of ICP-MS with a quadrupole analyzer to around 1%, which is insufficient for most
383:
A major issue that arises in SIMS analysis is the generation of isobaric interference between sputtered molecular ions and the ions of interest. This issue occurs with U–Pb dating as Pb ions have essentially the same mass as
358: 192:, an accurate determination of the isotopic make up of the sample is obtained. For example, carbon isotope ratios are measured relative to the international standard for C. The C standard is produced from a fossil 836:
Hinton, R.W. and Long, J.V.P. (1979). High-resolution ion microprobe measurement of lead isotopes: variations within single zircons from Lac Seul, Northwestern Ontario. Earth Planet. Sci. lett. 45, 309-325.,
404:
An MC-ICP-MS instrument is a multiple collector mass spectrometer with a plasma source. MC-ICP-MS was developed to improve the precision achievable by ICP-MS during isotope-ratio measurements. Conventional
124:
The isotope-ratio mass spectrometer (IRMS) allows the precise measurement of mixtures of naturally occurring isotopes. Most instruments used for precise determination of isotope ratios are of the magnetic
524:. The solution (or outflow from the chromatography) is dried onto a nickel or stainless steel wire. After the residue is deposited on the wire, it enters a furnace where the sample is converted to CO 188:
Samples must be introduced to the mass spectrometer as pure gases, achieved through combustion, gas chromatographic feeds, or chemical trapping. By comparing the detected isotopic ratios to a measured
360: 359: 1162: 251:
The two most common types of IRMS instruments are continuous flow and dual inlet. In dual inlet IRMS, purified gas obtained from a sample is alternated rapidly with a standard gas (of
356: 357: 380:
This assembly allows the secondary ions to be focused based on their kinetic energy and mass-charge ratio in order to be accurately collected using a series of Faraday cups.
344:(Hf-W). Though the TIMS method can create molecular ions instead in this case, species with high ionization potential can be analyzed more effectively with MC-ICP-MS. 1345: 1340: 1092: 252: 209: 205: 189: 620:
Stellaard F, Elzinga H (2005). "Analytical techniques in biomedical stable isotope applications: (isotope ratio) mass spectrometry or infrared spectrometry?".
1197: 1147: 1335: 1157: 1008: 849:
Caimi, R. J.; Brenna, J. T. (1996). "Direct analysis of carbon isotope variability in albumins by liquid flow-injection isotope ratio mass spectrometry".
1172: 1446: 1363: 1253: 1330: 1072: 1398: 508:
can distinguish between species with the same mass/charge ratio. Together, these processes allow the analysis of extreme isotope ratios above 10.
1388: 1152: 146:(m/z). Beams with lighter ions bend at a smaller radius than beams with heavier ions. The current of each ion beam is then measured using a ' 1077: 968: 937: 1082: 297: 1315: 1112: 1102: 1054: 577:
Paul D, Skrzypek G, FĂłrizs I (2007). "Normalization of measured stable isotopic compositions to isotope reference scales - a review".
1393: 1355: 1177: 783: 1378: 1233: 1142: 1107: 1182: 1001: 373: 101:
isotope analysis involves measuring the abundances of decay-products of natural radioactivity, and is used in most long-lived
1383: 1368: 174:, causing measurable effects on the isotopic composition of samples, characteristic of their biological or physical history. 313: 301: 372:
An alternative approach used to measure the relative abundance of radiogenic isotopes when working with a solid surface is
1373: 1320: 501: 1294: 1403: 1097: 542: 1434: 1279: 659: 166:
Measurement of natural variations in the abundances of stable isotopes of the same element is normally referred to as
1473: 1422: 1049: 994: 762:
Brenna JT, Corso TN, Tobias HJ, Caimi RJ (1997). "High-precision continuous-flow isotope ratio mass spectrometry".
665: 885: 888:(2005). "Moving-wire device for carbon isotopic analyses of nanogram quantities of nonvolatile organic carbon". 1325: 1310: 1238: 1223: 1028: 130: 305: 1122: 277: 208:(Vienna Pee Dee Belemnite) and has C:C ratio of 0.0112372. Oxygen isotope ratios are measured relative the 1192: 698: 94: 37: 1468: 1127: 521: 171: 170:
analysis. This field is of interest because the differences in mass between different isotopes leads to
89:
This technique has two different applications in the earth and environmental sciences. The analysis of '
441: 309: 727:
Meier-Augenstein, W. (1999). "Applied gas chromatography coupled to isotope ratio mass spectrometry".
771: 712:
C. B. Bouthitt; K. Garnett. "The Evolution of the Multicollector in Isotope Ratio Mass Spectromety".
586: 492: 273: 90: 1274: 1243: 1117: 1087: 1202: 418: 293: 289: 269: 102: 43: 361: 17: 1017: 974: 964: 943: 933: 905: 866: 787: 744: 637: 602: 557: 463: 142: 126: 113: 79: 63: 1187: 1167: 1132: 897: 858: 779: 736: 629: 594: 473: 197: 775: 590: 1137: 241: 220: 201: 167: 93:' is normally concerned with measuring isotopic variations arising from mass-dependent 740: 1462: 1284: 1044: 862: 827:
Dickin, A. P. (2005). Radiogenic Isotope Geology 2nd ed. Cambridge University Press.
82:, in which mass spectrometric methods are used to measure the relative abundance of 1248: 388:. In order to overcome this problem, a sensitive high-resolution ion microprobe ( 1289: 1228: 1207: 530: 477: 265: 155: 147: 138: 928:
Goetz, A.; Platzner, I. T. (Itzhak Thomas); Habfast, K.; Walder, A. J. (1997).
215: 1064: 633: 552: 411: 98: 978: 960: 947: 517: 325: 248:)) is purified by means of traps, filters, catalysts and/or chromatography. 193: 182: 909: 870: 748: 669: 641: 606: 791: 368:
instrument illustrating the ion beam path. After Figure 4, Williams, 1998.
42:
Magnetic sector mass spectrometer used in isotope ratio analysis, through
1258: 547: 341: 296:
of a solid sample loaded into the source of the mass spectrometer (hence
178: 446:
Accelerator mass spectrometer at Lawrence Livermore National Laboratory
83: 901: 784:
10.1002/(SICI)1098-2787(1997)16:5<227::AID-MAS1>3.0.CO;2-J
598: 406: 389: 365: 337: 986: 351: 214: 112: 520:
ratios of compounds in a solution, such as after purification by
329: 158:, which allows the simultaneous detection of multiple isotopes. 117:
Schematic of an isotope-ratio mass spectrometer for measuring CO
990: 400:
Multiple collector inductively coupled plasma mass spectrometry
1033: 808:
2nd ed. (Cambridge: Cambridge University Press), pp. 21-22.
200:, which is a limestone formed in the Cretaceous period in 27:
Usage of mass spectrometry to measure remaining isotopes
1354: 1303: 1267: 1216: 1063: 653: 651: 488: 483: 469: 459: 451: 59: 51: 533:of carbon can yield precise (within 1‰) results. 955:Yamasaki, Shin-ichi; Boutton, Thomas W. (1996). 500:For isotopes occurring at extremely low levels, 219:Isotope-ratio mass spectrometer used to measure 355: 129:type. This type of analyzer is superior to the 844: 842: 264:particularly used in the isotopic analysis of 1002: 212:, V-SMOW (Vienna Standard Mean Ocean Water). 8: 622:Isotopes in Environmental and Health Studies 434: 30: 288:Several of the isotope systems involved in 1009: 995: 987: 440: 36: 714:Proceedings of the 18th AMZSMS Conference 516:Moving wire IRMS is useful for analyzing 569: 137:The sector-type instrument designed by 97:in natural systems. On the other hand, 930:Modern isotope ratio mass spectrometry 433: 204:, U.S.A. The fossil is referred to as 29: 7: 1429: 692:Encyclopaedia of Analytical Science 298:thermal ionization mass spectrometry 284:Thermal ionization mass spectrometry 223:ratios, with gas bench in foreground 1441: 694:Encyclopaedia of Analytical Science 25: 1440: 1428: 1417: 1416: 374:secondary-ion mass spectrometry 348:Secondary-ion mass spectrometry 300:, TIMS). These methods include 72:Isotope-ratio mass spectrometry 31:Isotope-ratio mass spectrometry 18:Isotope ratio mass spectrometry 529:containing as little as 1 nano 1: 884:Sessions, A.L.; Sylva, S.P.; 741:10.1016/S0021-9673(98)01057-7 502:accelerator mass spectrometry 435:Accelerator mass spectrometry 430:Accelerator mass spectrometry 280:and helium isotope analysis. 863:10.1016/1044-0305(96)00010-4 543:Bainbridge mass spectrometer 259:Static gas mass spectrometry 162:Gas source mass spectrometry 1280:Microchannel plate detector 579:Rapid Commun. Mass Spectrom 417:Isotope-ratio analysis for 177:As a specific example, the 1490: 957:Mass spectrometry of soils 806:Radiogenic Isotope Geology 690:Townsend, A., ed. (1995). 666:Cambridge University Press 661:Radiogenic Isotope Geology 268:(rare or inert gases) for 253:known isotopic composition 150:' or multiplier detector. 1412: 1024: 851:J. Am. Soc. Mass Spectrom 764:Mass Spectrometry Reviews 634:10.1080/10256010500384333 439: 314:samarium–neodymium dating 302:rubidium–strontium dating 276:. Important examples are 78:) is a specialization of 35: 1295:Langmuir–Taylor detector 364:Schematic diagram of a 141:was such an advance in 1239:Quadrupole mass filter 699:Academic Press Limited 369: 224: 121: 95:isotopic fractionation 658:Dickin, A.P. (2005). 522:liquid chromatography 363: 292:depend on IRMS using 236:), carbon dioxide (CO 218: 172:isotope fractionation 116: 932:. London: J. Wiley. 890:Analytical Chemistry 804:Dickin, A.P., 2005. 493:Particle accelerator 274:isotope geochemistry 1275:Electron multiplier 1244:Quadrupole ion trap 776:1997MSRv...16..227B 591:2007RCMS...21.3006P 436: 306:uranium–lead dating 86:in a given sample. 32: 419:radiometric dating 370: 294:thermal ionization 290:radiometric dating 278:argon–argon dating 270:radiometric dating 225: 122: 103:radiometric dating 44:thermal ionization 1474:Mass spectrometry 1456: 1455: 1018:Mass spectrometry 970:978-0-8247-9699-0 939:978-0-471-97416-1 902:10.1021/ac051251z 896:(20): 6519–6527. 558:Table of nuclides 498: 497: 474:Organic molecules 464:Mass spectrometry 414:isotope systems. 143:mass spectrometer 80:mass spectrometry 69: 68: 64:mass spectrometry 16:(Redirected from 1481: 1444: 1443: 1432: 1431: 1420: 1419: 1011: 1004: 997: 988: 982: 951: 914: 913: 881: 875: 874: 846: 837: 834: 828: 825: 819: 815: 809: 802: 796: 795: 759: 753: 752: 735:(1–2): 351–371. 729:J. Chromatogr. A 724: 718: 717: 709: 703: 702: 687: 681: 680: 678: 677: 668:. Archived from 655: 646: 645: 617: 611: 610: 599:10.1002/rcm.3185 574: 512:Moving wire IRMS 484:Other techniques 444: 437: 354: 310:lead–lead dating 198:Peedee Formation 40: 33: 21: 1489: 1488: 1484: 1483: 1482: 1480: 1479: 1478: 1459: 1458: 1457: 1452: 1408: 1350: 1299: 1263: 1212: 1059: 1020: 1015: 985: 971: 954: 940: 927: 923: 918: 917: 883: 882: 878: 848: 847: 840: 835: 831: 826: 822: 816: 812: 803: 799: 761: 760: 756: 726: 725: 721: 711: 710: 706: 689: 688: 684: 675: 673: 657: 656: 649: 619: 618: 614: 585:(18): 3006–14. 576: 575: 571: 566: 539: 527: 514: 476: 447: 432: 402: 395: 387: 362: 352: 350: 286: 261: 247: 239: 235: 231: 164: 120: 111: 91:stable isotopes 47: 28: 23: 22: 15: 12: 11: 5: 1487: 1485: 1477: 1476: 1471: 1461: 1460: 1454: 1453: 1451: 1450: 1438: 1426: 1413: 1410: 1409: 1407: 1406: 1401: 1396: 1391: 1386: 1381: 1376: 1371: 1366: 1360: 1358: 1352: 1351: 1349: 1348: 1343: 1338: 1333: 1328: 1323: 1318: 1313: 1307: 1305: 1304:MS combination 1301: 1300: 1298: 1297: 1292: 1287: 1282: 1277: 1271: 1269: 1265: 1264: 1262: 1261: 1256: 1251: 1246: 1241: 1236: 1234:Time-of-flight 1231: 1226: 1220: 1218: 1214: 1213: 1211: 1210: 1205: 1200: 1195: 1190: 1185: 1180: 1175: 1170: 1165: 1160: 1155: 1150: 1145: 1140: 1135: 1130: 1125: 1120: 1115: 1110: 1105: 1100: 1095: 1090: 1085: 1080: 1075: 1069: 1067: 1061: 1060: 1058: 1057: 1052: 1047: 1042: 1031: 1025: 1022: 1021: 1016: 1014: 1013: 1006: 999: 991: 984: 983: 969: 952: 938: 924: 922: 919: 916: 915: 876: 857:(6): 605–610. 838: 829: 820: 810: 797: 754: 719: 704: 682: 647: 612: 568: 567: 565: 562: 561: 560: 555: 550: 545: 538: 535: 525: 513: 510: 496: 495: 490: 486: 485: 481: 480: 471: 467: 466: 461: 460:Classification 457: 456: 453: 449: 448: 445: 431: 428: 401: 398: 393: 385: 349: 346: 285: 282: 260: 257: 245: 242:sulfur dioxide 237: 233: 232:), nitrogen (N 229: 221:stable isotope 202:South Carolina 168:stable isotope 163: 160: 118: 110: 107: 67: 66: 61: 60:Classification 57: 56: 53: 49: 48: 41: 26: 24: 14: 13: 10: 9: 6: 4: 3: 2: 1486: 1475: 1472: 1470: 1467: 1466: 1464: 1449: 1448: 1439: 1437: 1436: 1427: 1425: 1424: 1415: 1414: 1411: 1405: 1402: 1400: 1397: 1395: 1392: 1390: 1387: 1385: 1382: 1380: 1377: 1375: 1372: 1370: 1367: 1365: 1362: 1361: 1359: 1357: 1356:Fragmentation 1353: 1347: 1344: 1342: 1339: 1337: 1334: 1332: 1329: 1327: 1324: 1322: 1319: 1317: 1314: 1312: 1309: 1308: 1306: 1302: 1296: 1293: 1291: 1288: 1286: 1285:Daly detector 1283: 1281: 1278: 1276: 1273: 1272: 1270: 1266: 1260: 1257: 1255: 1252: 1250: 1247: 1245: 1242: 1240: 1237: 1235: 1232: 1230: 1227: 1225: 1222: 1221: 1219: 1217:Mass analyzer 1215: 1209: 1206: 1204: 1201: 1199: 1196: 1194: 1191: 1189: 1186: 1184: 1181: 1179: 1176: 1174: 1171: 1169: 1166: 1164: 1161: 1159: 1156: 1154: 1151: 1149: 1146: 1144: 1141: 1139: 1136: 1134: 1131: 1129: 1126: 1124: 1121: 1119: 1116: 1114: 1111: 1109: 1106: 1104: 1101: 1099: 1096: 1094: 1091: 1089: 1086: 1084: 1081: 1079: 1076: 1074: 1071: 1070: 1068: 1066: 1062: 1056: 1053: 1051: 1048: 1046: 1045:Mass spectrum 1043: 1041: 1040: 1036: 1032: 1030: 1027: 1026: 1023: 1019: 1012: 1007: 1005: 1000: 998: 993: 992: 989: 980: 976: 972: 966: 962: 958: 953: 949: 945: 941: 935: 931: 926: 925: 920: 911: 907: 903: 899: 895: 891: 887: 880: 877: 872: 868: 864: 860: 856: 852: 845: 843: 839: 833: 830: 824: 821: 814: 811: 807: 801: 798: 793: 789: 785: 781: 777: 773: 770:(5): 227–58. 769: 765: 758: 755: 750: 746: 742: 738: 734: 730: 723: 720: 715: 708: 705: 700: 696: 693: 686: 683: 672:on 2014-03-27 671: 667: 663: 662: 654: 652: 648: 643: 639: 635: 631: 628:(4): 345–61. 627: 623: 616: 613: 608: 604: 600: 596: 592: 588: 584: 580: 573: 570: 563: 559: 556: 554: 551: 549: 546: 544: 541: 540: 536: 534: 532: 523: 519: 511: 509: 505: 503: 494: 491: 487: 482: 479: 475: 472: 468: 465: 462: 458: 454: 450: 443: 438: 429: 427: 423: 420: 415: 413: 408: 399: 397: 391: 381: 377: 375: 367: 347: 345: 343: 339: 333: 331: 327: 321: 317: 315: 311: 307: 303: 299: 295: 291: 283: 281: 279: 275: 271: 267: 258: 256: 254: 249: 243: 222: 217: 213: 211: 207: 203: 199: 196:found in the 195: 191: 186: 184: 180: 175: 173: 169: 161: 159: 157: 151: 149: 144: 140: 135: 132: 128: 115: 108: 106: 104: 100: 96: 92: 87: 85: 81: 77: 73: 65: 62: 58: 54: 50: 45: 39: 34: 19: 1469:Geochemistry 1445: 1433: 1421: 1249:Penning trap 1038: 1034: 959:. New York: 956: 929: 921:Bibliography 893: 889: 879: 854: 850: 832: 823: 813: 805: 800: 767: 763: 757: 732: 728: 722: 713: 707: 695: 691: 685: 674:. Retrieved 670:the original 660: 625: 621: 615: 582: 578: 572: 515: 506: 499: 478:Biomolecules 424: 416: 403: 382: 378: 371: 334: 322: 318: 287: 262: 250: 226: 187: 176: 165: 156:Faraday cups 152: 136: 123: 109:Introduction 88: 75: 71: 70: 1447:WikiProject 1290:Faraday cup 1229:Wien filter 1050:MS software 886:Hayes, J.M. 266:noble gases 148:Faraday cup 139:Alfred Nier 1463:Categories 1065:Ion source 697:. London: 676:2008-10-09 564:References 553:Isotopomer 412:radiogenic 340:(Os), and 328:(Sr), and 131:quadrupole 99:radiogenic 1326:Hybrid MS 961:M. Dekker 716:: THO–07. 518:carbon-13 326:strontium 194:belemnite 183:deuterium 105:methods. 1423:Category 1268:Detector 1259:Orbitrap 1055:Acronyms 979:34473560 948:36461690 910:16223235 871:24203433 749:10377971 642:16543190 607:17705258 548:Isoscape 537:See also 470:Analytes 342:tungsten 210:standard 190:standard 181:isotope 179:hydrogen 84:isotopes 1435:Commons 1163:MALDESI 792:9538528 772:Bibcode 587:Bibcode 489:Related 452:Acronym 52:Acronym 1341:IMS/MS 1254:FT-ICR 1224:Sector 977:  967:  946:  936:  908:  869:  790:  747:  640:  605:  407:ICP-MS 390:SHRIMP 366:SHRIMP 338:osmium 332:(Pb). 240:), or 127:sector 1394:IRMPD 1346:CE-MS 1336:LC/MS 1331:GC/MS 1311:MS/MS 1198:SELDI 1158:MALDI 1153:LAESI 1093:DAPPI 1399:NETD 1364:BIRD 1183:SIMS 1178:SESI 1113:EESI 1108:DIOS 1103:DESI 1098:DART 1083:APPI 1078:APLI 1073:APCI 1029:Mass 975:OCLC 965:ISBN 944:OCLC 934:ISBN 906:PMID 867:PMID 818:1–35 788:PMID 745:PMID 638:PMID 603:PMID 531:mole 330:lead 312:and 206:VPDB 76:IRMS 55:IRMS 1404:SID 1389:HCD 1384:ETD 1379:EDD 1374:ECD 1369:CID 1321:AMS 1316:QqQ 1193:SSI 1173:PTR 1168:MIP 1148:ICP 1128:FAB 1123:ESI 898:doi 859:doi 780:doi 737:doi 733:842 630:doi 595:doi 455:AMS 384:HfO 272:or 244:(SO 1465:: 1208:TS 1203:TI 1188:SS 1143:IA 1138:GD 1133:FD 1118:EI 1088:CI 973:. 963:. 942:. 904:. 894:77 892:. 865:. 853:. 841:^ 786:. 778:. 768:16 766:. 743:. 731:. 664:. 650:^ 636:. 626:41 624:. 601:. 593:. 583:21 581:. 396:. 316:. 308:, 304:, 228:(H 1039:z 1037:/ 1035:m 1010:e 1003:t 996:v 981:. 950:. 912:. 900:: 873:. 861:: 855:7 794:. 782:: 774:: 751:. 739:: 701:. 679:. 644:. 632:: 609:. 597:: 589:: 526:2 394:2 386:2 246:2 238:2 234:2 230:2 119:2 74:( 46:. 20:)

Index

Isotope ratio mass spectrometry

thermal ionization
mass spectrometry
mass spectrometry
isotopes
stable isotopes
isotopic fractionation
radiogenic
radiometric dating

sector
quadrupole
Alfred Nier
mass spectrometer
Faraday cup
Faraday cups
stable isotope
isotope fractionation
hydrogen
deuterium
standard
belemnite
Peedee Formation
South Carolina
VPDB
standard

stable isotope
sulfur dioxide

Text is available under the Creative Commons Attribution-ShareAlike License. Additional terms may apply.

↑