Knowledge (XXG)

Size-exclusion chromatography

Source 📝

841:
measurement; for example, with MALS analysis, the molar mass of inherently disordered proteins are characterized accurately even though they elute at much earlier times than globular proteins with the same molar mass, and the same is true of branched polymers which elute late compared to linear reference standards with the same molar mass. Another benefit of ASEC is that the molar mass and/or size is determined at each point in an eluting peak, and therefore indicates homogeneity or polydispersity within the peak. For example, SEC-MALS analysis of a monodisperse protein will show that the entire peak consists of molecules with the same molar mass, something that is not possible with standard SEC analysis.
405:("stationary phase"). This process is usually performed within a column, which typically consists of a hollow tube tightly packed with micron-scale polymer beads containing pores of different sizes. These pores may be depressions on the surface or channels through the bead. As the solution travels down the column some particles enter into the pores. Larger particles cannot enter into as many pores. The larger the particles, the faster the elution. The larger molecules simply pass by the pores because those molecules are too large to enter the pores. Larger molecules therefore flow through the column more quickly than smaller molecules, that is, the smaller the molecule, the longer the retention time. 429:
and computer simulations assume a thermodynamic separation principle: the separation process is determined by the equilibrium distribution (partitioning) of solute macromolecules between two phases: a dilute bulk solution phase located at the interstitial space and confined solution phases within the pores of column packing material. Based on this theory, it has been shown that the relevant size parameter to the partitioning of polymers in pores is the mean span dimension (mean maximal projection onto a line). Although this issue has not been fully resolved, it is likely that the mean span dimension and the hydrodynamic volume are strongly correlated.
409:
phases. Thus, a small molecule that can penetrate every region of the stationary phase pore system can enter a total volume equal to the sum of the entire pore volume and the interparticle volume. This small molecule elutes late (after the molecule has penetrated all of the pore- and interparticle volume—approximately 80% of the column volume). At the other extreme, a very large molecule that cannot penetrate any the smaller pores can enter only the interparticle volume (~35% of the column volume) and elutes earlier when this volume of mobile phase has passed through the column. The underlying principle of SEC is that particles of different sizes
387: 433: 688: 803:
weight and hydrodynamic volume for polystyrene can be found. For this, polystyrene is used as a standard. But the relationship between hydrodynamic volume and molecular weight is not the same for all polymers, so only an approximate measurement can be obtained. Another drawback is the possibility of interaction between the stationary phase and the analyte. Any interaction leads to a later elution time and thus mimics a smaller analyte size.
696: 38: 2455: 865:
allowing oligomeric resolution. Aggregation studies can also be done using ASEC. Though the aggregate concentration may not be calculated with light scattering (an online concentration detector such as that used in SEC-MALS for molar mass measurement also determines aggregate concentration), the size of the aggregate can be measured, only limited by the maximum size eluting from the SEC columns.
225:
technique is generally combined with others that further separate molecules by other characteristics, such as acidity, basicity, charge, and affinity for certain compounds. With size exclusion chromatography, there are short and well-defined separation times and narrow bands, which lead to good sensitivity. There is also no sample loss because solutes do not interact with the stationary phase.
212:, where an electric field is used to "pull" molecules through the gel depending on their electrical charges. The amount of time a solute remains within a pore is dependent on the size of the pore. Larger solutes will have access to a smaller volume and vice versa. Therefore, a smaller solute will remain within the pore for a longer period of time compared to a larger solute. 675:
An under-packed column can reduce the relative surface area of the stationary phase accessible to smaller species, resulting in those species spending less time trapped in pores. Unlike affinity chromatography techniques, a solvent head at the top of the column can drastically diminish resolution as the sample diffuses prior to loading, broadening the downstream elution.
869:
measurement time is essentially instantaneous, and the range of concentration is several orders of magnitude larger than for DLS. However, molar mass analysis with SEC-MALS does require accurate concentration measurements. MALS and DLS detectors are often combined in a single instrument for more comprehensive absolute analysis following separation by SEC.
656: 812:
bands broadening. For instance, one can apply the sample in a narrow, highly concentrated band on the top of the column. The more concentrated the eluent is, the more efficient the procedure would be. However, it is not always possible to concentrate the eluent, which can be considered as one more disadvantage.
860:
In SEC-DLS, the sizes of the macromolecules are measured as they elute into the flow cell of the DLS instrument from the size exclusion column set. The hydrodynamic size of the molecules or particles are measured and not their molecular weights. For proteins a Mark-Houwink type of calculation can be
674:
Like other forms of chromatography, increasing the column length enhances resolution, and increasing the column diameter increases column capacity. Proper column packing is important for maximum resolution: An over-packed column can collapse the pores in the beads, resulting in a loss of resolution.
440:
Each size exclusion column has a range of molecular weights that can be separated. The exclusion limit defines the molecular weight at the upper end of the column 'working' range and is where molecules are too large to get trapped in the stationary phase. The lower end of the range is defined by the
341:
component represents the volume at which the larger molecules elute, which elute in the beginning. Disadvantages are, for example, that only a limited number of bands can be accommodated because the time scale of the chromatogram is short, and, in general, there must be a 10% difference in molecular
224:
The advantages of this method include good separation of large molecules from the small molecules with a minimal volume of eluate, and that various solutions can be applied without interfering with the filtration process, all while preserving the biological activity of the particles to separate. The
864:
A major advantage of DLS coupled with SEC is the ability to obtain enhanced DLS resolution. Batch DLS is quick and simple and provides a direct measure of the average size, but the baseline resolution of DLS is a ratio of 3:1 in diameter. Using SEC, the proteins and protein oligomers are separated,
802:
In SEC, mass is not measured so much as the hydrodynamic volume of the polymer molecules, that is, how much space a particular polymer molecule takes up when it is in solution. However, the approximate molecular weight can be calculated from SEC data because the exact relationship between molecular
789:
can be used online with SEC to yield absolute molecular weights that do not rely on calibration with standards of known molecular weight. Due to the difference in size of two polymers with identical molecular weights, the absolute determination methods are, in general, more desirable. A typical SEC
428:
Still, the use of the hydrodynamic volume, a size based on dynamical properties, in the interpretation of SEC data is not fully understood. This is because SEC is typically run under low flow rate conditions where hydrodynamic factor should have little effect on the separation. In fact, both theory
811:
between the molecules of the glass walls and the molecules of the eluent contribute to the broadening of the bands. Besides broadening, the bands also overlap with each other. As a result, the eluent usually gets considerably diluted. A few precautions can be taken to prevent the likelihood of the
420:
and the hydrodynamic radius), a fundamental problem in the theory of SEC has been the choice of a proper molecular size parameter by which molecules of different kinds are separated. Experimentally, Benoit and co-workers found an excellent correlation between elution volume and a dynamically based
228:
The other advantage to this experimental method is that in certain cases, it is feasible to determine the approximate molecular weight of a compound. The shape and size of the compound (eluent) determine how the compound interacts with the gel (stationary phase). To determine approximate molecular
215:
Even though size exclusion chromatography is widely utilized to study natural organic material, there are limitations. One of these limitations include that there is no standard molecular weight marker; thus, there is nothing to compare the results back to. If precise molecular weight is required,
408:
One requirement for SEC is that the analyte does not interact with the surface of the stationary phases, with differences in elution time between analytes ideally being based solely on the solute volume the analytes can enter, rather than chemical or electrostatic interactions with the stationary
868:
Limitations of ASEC with DLS detection include flow-rate, concentration, and precision. Because a correlation function requires anywhere from 3–7 seconds to properly build, a limited number of data points can be collected across the peak. ASEC with SLS detection is not limited by flow rate and
840:
in units of nm). Non-ideal column interactions, such as electrostatic or hydrophobic surface interactions that modulate retention time relative to standards, do not impact the final result. Likewise, differences between conformation of the analyte and the standard have no effect on an absolute
666:
In real-life situations, particles in solution do not have a fixed size, resulting in the probability that a particle that would otherwise be hampered by a pore passing right by it. Also, the stationary-phase particles are not ideally defined; both particles and pores may vary in size. Elution
683:
In simple manual columns, the eluent is collected in constant volumes, known as fractions. The more similar the particles are in size the more likely they are in the same fraction and not detected separately. More advanced columns overcome this problem by constantly monitoring the eluent.
734:. (Blue dextran is not recommended for Vo determination because it is heterogeneous and may give variable results) The elution volumes of the standards are divided by the elution volume of the thyroglobulin (Ve/Vo) and plotted against the log of the standards' molecular weights. 711:(RI) and ultraviolet (UV). When eluting spectroscopically similar species (such as during biological purification), other techniques may be necessary to identify the contents of each fraction. It is also possible to analyze the eluent flow continuously with RI, 413:(filter) through a stationary phase at different rates. This results in the separation of a solution of particles based on size. Provided that all the particles are loaded simultaneously or near-simultaneously, particles of the same size should elute together. 806:
When performing this method, the bands of the eluting molecules may be broadened. This can occur by turbulence caused by the flow of the mobile phase molecules passing through the molecules of the stationary phase. In addition, molecular thermal diffusion and
350:
The technique was invented in 1955 by Grant Henry Lathe and Colin R Ruthven, working at Queen Charlotte's Hospital, London. They later received the John Scott Award for this invention. While Lathe and Ruthven used starch gels as the matrix,
835:
The definition of “absolute” in this case is that calibration of retention time on the column with a set of reference standards is not required to obtain molar mass or the hydrodynamic size, often referred to as hydrodynamic diameter
671:. The stationary phase may also interact in undesirable ways with a particle and influence retention times, though great care is taken by column manufacturers to use stationary phases that are inert and minimize this issue. 441:
permeation limit, which defines the molecular weight of a molecule that is small enough to penetrate all pores of the stationary phase. All molecules below this molecular mass are so small that they elute as a single band.
194:
Size exclusion chromatography (SEC) is fundamentally different from all other chromatographic techniques in that separation is based on a simple procedure of classifying molecule sizes rather than any type of interaction.
747:
In general, SEC is considered a low-resolution chromatography as it does not discern similar species very well, and is therefore often reserved for the final step of a purification. The technique can determine the
207:
of proteins and other water-soluble polymers, while gel permeation chromatography is used to analyze the molecular weight distribution of organic-soluble polymers. Either technique should not be confused with
760:
of a typical protein domain might be 14 Å and 36 Å for the folded and unfolded forms, respectively. SEC allows the separation of these two forms, as the folded form elutes much later due to its smaller size.
844:
Determination of molar mass with SLS requires combining the light scattering measurements with concentration measurements. Therefore SEC-MALS typically includes the light scattering detector and either a
374:
with controlled pore size, that a rapid increase of research activity in this field began. It was recognized almost immediately that with proper calibration, GPC was capable to provide molar mass and
730:. Columns are often calibrated using 4-5 standard samples (e.g., folded proteins of known molecular weight), and a sample containing a very large molecule such as thyroglobulin to determine the 327: 425:, for several different chain architecture and chemical compositions. The observed correlation based on the hydrodynamic volume became accepted as the basis of universal SEC calibration. 832:(DLS) instrument, to a size-exclusion chromatography system for absolute molar mass and/or size measurements of proteins and macromolecules as they elute from the chromatography system. 2372: 752:
of purified proteins that have slow exchange times, since it can be carried out under native solution conditions, preserving macromolecular interactions. SEC can also assay protein
944:
Paul-Dauphin, S; Karaca, F; Morgan, TJ; et al. (6 Oct 2007). "Probing Size Exclusion Mechanisms of Complex Hydrocarbon Mixtures: The Effect of Altering Eluent Compositions".
857:. For smaller molecules, either DLS or, more commonly, a differential viscometer is added to determine hydrodynamic radius and evaluate molecular conformation in the same manner. 1946: 1162: 756:, as it measures the hydrodynamic volume (not molecular weight), allowing folded and unfolded versions of the same protein to be distinguished. For example, the apparent 2118: 2087: 2031: 1979: 359:
later introduced dextran gels; other gels with size fractionation properties include agarose and polyacrylamide. A short review of these developments has appeared.
853:
of molecules above a certain size limit, typically 10 nm. SEC-MALS can therefore analyze the conformation of polymers via the relationship of molar mass to R
1119:
Müller MB, Schmitt D, Frimmel FH (1 Dec 2000). "Fractionation of Natural Organic Matter by Size Exclusion Chromatography−Properties and Stability of Fractions".
790:
system can quickly (in about half an hour) give polymer chemists information on the size and polydispersity of the sample. The preparative SEC can be used for
333:
component represents the volume at which the intermediate molecules elute such as molecules that have partial access to the beads of the column. In addition, V
167:, which is used when an organic solvent is used as a mobile phase. The chromatography column is packed with fine, porous beads which are commonly composed of 2367: 1939: 2362: 2230: 401:
SEC is used primarily for the analysis of large molecules such as proteins or polymers. SEC works by trapping smaller molecules in the pores of the
2015: 378:
information for synthetic polymers. Because the latter information was difficult to obtain by other methods, GPC came rapidly into extensive use.
329:
and Mw is the molecular mass. This plot acts as a calibration curve, which is used to approximate the desired compound's molecular weight. The V
2321: 2316: 2301: 100: 2489: 1709: 1121: 1036: 1932: 1897: 777:, that is, the ability to find the distribution of the sizes of polymer molecules. If standards of a known size are run previously, then a 1830:"Truths and myths about the determination of molar mass distribution of synthetic and natural polymers by size exclusion chromatography" 104: 1655:
Wang Y, Teraoka I, Hansen FY, et al. (2010). "A Theoretical Study of the Separation Principle in Size Exclusion Chromatography".
1226: 716: 394: 1870: 1760: 1567: 1541: 1497: 1262: 1177: 920: 2041: 712: 2134: 2306: 1989: 2484: 2346: 112: 2494: 2331: 2077: 883: 367: 164: 1610:
Sun T, Chance RR, Graessley WW, Lohse DJ (2004). "A Study of the Separation Principle in Size Exclusion Chromatography".
362:
There were also attempts to fractionate synthetic high polymers; however, it was not until 1964, when J. C. Moore of the
2223: 2198: 2176: 1321:"The separation of substances and estimation of their relative molecular sizes by the use of columns of starch in water" 821: 236: 1916: 820:
Absolute size-exclusion chromatography (ASEC) is a technique that couples a light scattering instrument, most commonly
2439: 2432: 2286: 1495:
Moore JC (1964). "Gel permeation chromatography. I. A new method for molecular weight distribution of high polymers".
229:
weight, the elution volumes of compounds with their corresponding molecular weights are obtained and then a plot of “K
108: 2276: 2193: 846: 1756:"Characterization of Proteins by Size-Exclusion Chromatography Coupled to Multi-Angle Light Scattering (SEC-MALS)" 2425: 2341: 2255: 2139: 781:
can be created to determine the sizes of polymer molecules of interest in the solvent chosen for analysis (often
2082: 1984: 1969: 1657: 1612: 829: 184: 386: 2479: 2458: 2271: 2250: 2216: 2103: 2056: 825: 375: 188: 459:
Following are the materials which are commonly used for porous gel beads in size exclusion chromatography
2398: 432: 1798: 1534:
Modern Size-Exclusion Liquid Chromatography: Practice of Gel Permeation and Gel Filtration Chromatography
1280:"The separation of substances on the basis of their molecular weights, using columns of starch and water" 2388: 2291: 2281: 2160: 2144: 2113: 1974: 791: 699:
Size exclusion chromatogram after bioanalytical continuous-elution gel chromatography of a plant sample
27:
Chromatographic method in which dissolved molecules are separated by their size & molecular weight
2504: 2408: 2326: 1666: 1621: 1576: 1402: 1130: 1065: 888: 749: 707:
to determine the concentration of the particles eluted. Common spectroscopy detection techniques are
363: 132: 42:
Equipment for running size-exclusion chromatography. The buffer is pumped through the column (left).
2393: 2181: 1565:
Grubisic Z, Rempp P, Benoit H (1967). "A universal calibration for gel permeation chromatography".
1218: 757: 727: 668: 422: 209: 1911: 687: 2403: 2311: 2296: 2108: 2036: 1477: 1426: 1083: 753: 731: 417: 136: 356: 1393:
Porath J, Flodin P (June 1959). "Gel filtration: a method for desalting and group separation".
2499: 2010: 2005: 1955: 1876: 1866: 1779: 1705: 1682: 1637: 1592: 1547: 1537: 1514: 1469: 1418: 1350: 1301: 1258: 1232: 1222: 1191: 1183: 1173: 1101: 1032: 1009: 946: 926: 916: 778: 73: 452:
includes any particles too large to enter the medium, and the solvent volume is known as the
17: 1841: 1810: 1769: 1674: 1629: 1584: 1506: 1459: 1410: 1368: 1340: 1332: 1291: 1210: 1138: 1091: 1073: 999: 989: 955: 708: 156: 140: 2072: 978:"Phytochemical Approach and Bioanalytical Strategy to Develop Chaperone-Based Medications" 1211: 416:
However, as there are various measures of the size of a macromolecule (for instance, the
1670: 1625: 1580: 1406: 1134: 1069: 2239: 1450: 1345: 1320: 1296: 1279: 1004: 977: 770: 337:
is the sum of the total volume between the beads and the volume within the beads. The V
176: 128: 59: 37: 2473: 1814: 1096: 1053: 352: 204: 180: 69: 1481: 1430: 2051: 704: 994: 371: 77: 1588: 1510: 1058:
Proceedings of the National Academy of Sciences of the United States of America
179:
polymers. The pore sizes of these beads are used to estimate the dimensions of
2046: 1924: 930: 878: 786: 159:
is used to transport the sample through the column, the technique is known as
1880: 1686: 1641: 1596: 1551: 1518: 1473: 1236: 1195: 849:
or UV/Vis absorbance detector. In addition, MALS determines the rms radius R
723: 695: 402: 1783: 1422: 1354: 1305: 1105: 1078: 1013: 1464: 1445: 915:(5th ed.). Belmont, CA: Brooks/Cole, Cengage Learning. p. 108. 808: 144: 655: 774: 410: 390: 172: 168: 152: 148: 1846: 1829: 1799:"Light scattering and the absolute characterization of macromolecules" 1678: 1633: 1414: 1336: 1142: 1087: 959: 785:). In alternative fashion, techniques such as light scattering and/or 1255:
Fundamental laboratory approaches for biochemistry and biotechnology
1865:. Vol. 2. Norfolk, UK: Horizon Scientific Press. p. 170. 444:
The filtered solution that is collected at the end is known as the
1774: 1755: 1187: 861:
used to estimate the molecular weight from the hydrodynamic size.
694: 686: 654: 431: 385: 2208: 1725: 2212: 1928: 1702:
Fundamentals and techniques of Biophysics and Molecular biology
1172:(6th ed.). Belmont, CA: Thomson Brooks/Cole. p. 816. 1863:
Lab on a chip technology: Biomolecular separation and analysis
782: 1213:
Chemical analysis: modern instrumental methods and techniques
203:
The main application of size-exclusion chromatography is the
393:-based SEC columns used for protein purification on an AKTA 87:
Cytiva, Bio-Rad, Bio-Works, emp Biotech, Knauer, Phenomenex.
722:
The elution volume (Ve) decreases roughly linear with the
1052:
Brooks DE, Haynes CA, Hritcu D, et al. (June 2000).
1257:(2nd ed.). Hoboken, N.J.: Wiley. pp. 127–129. 691:
Standardization (calibration) of a size exclusion column
1754:
Some, D; Amartely, H; Tsadok, A; Lebendiker, M (2019).
1054:"Size exclusion chromatography does not require pores" 769:
SEC can be used as a measure of both the size and the
1369:"The John Scott Award Recipients From 1822 - present" 239: 2417: 2381: 2355: 2264: 2169: 2153: 2127: 2096: 2065: 2024: 1998: 1962: 96: 91: 83: 65: 55: 47: 322:{\displaystyle K_{av}=(V_{e}-V_{o})/(V_{t}-V_{o})} 321: 1704:. New Delhi: Pathfinder Publication. p. 05. 1532:Striegel A, Yau WW, Kirkland JJ, Bly DD (2009). 1029:Practical high-performance liquid chromatography 1027:Meyer, Veronika R.; Meyer, Veronika R. (2010). 2373:Pyrolysis–gas chromatography–mass spectrometry 1732:. Polymer Science Learning Center (PSLC). 2005 1217:(Engl. ed.). Chichester: Wiley. pp.  906: 904: 703:The collected fractions are often examined by 187:method because of its ability to provide good 2224: 1940: 8: 30: 2231: 2217: 2209: 1947: 1933: 1925: 36: 1845: 1773: 1463: 1344: 1295: 1095: 1077: 1003: 993: 719:MALS, UV, and/or viscosity measurements. 659:A cartoon illustrating the theory behind 366:published his work on the preparation of 310: 297: 285: 276: 263: 244: 238: 461: 2368:Liquid chromatography–mass spectrometry 1161:Skoog DA, Holler FJ, Crouch SR (2006). 900: 2317:Micellar electrokinetic chromatography 2302:High-performance liquid chromatography 1749: 1747: 1536:(2nd ed.). Hoboken, N.J.: Wiley. 1253:Ballou DP, Benore M, Ninfa AJ (2008). 816:Absolute size-exclusion chromatography 101:High-performance liquid chromatography 29: 7: 2363:Gas chromatography–mass spectrometry 1797:Wyatt, Philip J. (1 February 1993). 1278:Lathe GH, Ruthven CR (August 1955). 1248: 1246: 1156: 1154: 1152: 971: 969: 370:(GPC) columns based on cross-linked 147:or macromolecular complexes such as 1319:Lathe GH, Ruthven CR (April 1956). 1170:Principles of instrumental analysis 105:Aqueous normal-phase chromatography 1834:Journal of Applied Polymer Science 1828:Podzimek, Stepan (April 5, 2014). 1031:(5. ed.). Chichester: Wiley. 717:Multi-Angle Laser Light Scattering 25: 1761:Journal of Visualized Experiments 143:. It is usually applied to large 2454: 2453: 2042:Dual-polarization interferometry 342:mass to have a good resolution. 2307:Capillary electrochromatography 1209:Rouessac A, Rouessac F (2000). 1163:"Ch. 28. Liquid Chromatography" 911:Garrett RH, Grisham CM (2013). 2347:Two-dimensional chromatography 2073:Analytical ultracentrifugation 1726:"Size Exclusion Chromatograhy" 316: 290: 282: 256: 233:” vs “log(Mw)” is made, where 216:other methods should be used. 125:molecular sieve chromatography 113:Micellar liquid chromatography 1: 2337:Size-exclusion chromatography 2332:Reversed-phase chromatography 2078:Size exclusion chromatography 1903:Gel permeation chromatography 1861:Herold KE, Rasooly A (2009). 982:The Open Biochemistry Journal 976:Kastenholz, B (29 Apr 2008). 884:Gel permeation chromatography 661:size exclusion chromatography 368:gel permeation chromatography 165:gel permeation chromatography 161:gel-filtration chromatography 131:method in which molecules in 121:Size-exclusion chromatography 31:Size-exclusion chromatography 18:Size exclusion chromatography 2490:Molecular biology techniques 2177:Protein structure prediction 1917:Resources in other libraries 1815:10.1016/0003-2670(93)80373-S 822:multi-angle light scattering 667:curves, therefore, resemble 651:Factors affecting filtration 2440:Journal of Chromatography B 2433:Journal of Chromatography A 2322:Normal-phase chromatography 2287:Displacement chromatography 2135:Hydrogen–deuterium exchange 1956:Protein structural analysis 995:10.2174/1874091X00802010044 191:(Mw) results for polymers. 109:Ion exchange chromatography 2521: 2277:Argentation chromatography 1589:10.1002/pol.1967.110050903 1511:10.1002/pol.1964.100020220 1446:"Adventures in the matrix" 1373:garfield.library.upenn.edu 847:differential refractometer 824:(MALS) or another form of 2449: 2426:Biomedical Chromatography 2342:Thin-layer chromatography 2246: 2190: 2140:Site-directed mutagenesis 1912:Resources in your library 35: 1985:Electron crystallography 1970:Cryo-electron microscopy 830:dynamic light scattering 794:on an analytical scale. 743:Biochemical applications 705:spectroscopic techniques 185:polymer characterization 2272:Affinity chromatography 2104:Fluorescence anisotropy 2066:Translational Diffusion 2057:Fluorescence anisotropy 1325:The Biochemical Journal 1284:The Biochemical Journal 826:static light scattering 436:A size exclusion column 376:molar mass distribution 189:molar mass distribution 183:. SEC is a widely used 135:are separated by their 2418:Prominent publications 2399:Kovats retention index 1803:Analytica Chimica Acta 1700:Kumar, Pranav (2018). 1079:10.1073/pnas.120129097 828:(SLS), but possibly a 700: 692: 669:Gaussian distributions 663: 437: 398: 323: 2485:Laboratory techniques 2389:Distribution constant 2292:Electrochromatography 2282:Column chromatography 2199:Quaternary structure→ 2161:Equilibrium unfolding 2145:Chemical modification 2114:Dielectric relaxation 1975:X-ray crystallography 1465:10.1038/nmeth0506-410 1444:Eisenstein M (2006). 792:polymer fractionation 698: 690: 658: 435: 389: 324: 155:. Typically, when an 2495:Biochemistry methods 2409:Van Deemter equation 2327:Paper chromatography 2097:Rotational Diffusion 889:Protein purification 750:quaternary structure 476:Fractionation range 421:molecular size, the 364:Dow Chemical Company 237: 139:, and in some cases 2394:Freundlich equation 2194:←Tertiary structure 1671:2010MaMol..43.1651W 1626:2004MaMol..37.4304S 1581:1967JPoSL...5..753G 1407:1959Natur.183.1657P 1135:2000EnST...34.4867M 1122:Environ Sci Technol 1070:2000PNAS...97.7064B 758:hydrodynamic radius 728:hydrodynamic volume 463: 423:hydrodynamic volume 210:gel electrophoresis 32: 2356:Hyphenated methods 2312:Ion chromatography 2297:Gas chromatography 2109:Flow birefringence 2037:Circular dichroism 947:Energy & Fuels 754:tertiary structure 701: 693: 664: 462: 438: 418:radius of gyration 399: 319: 163:, versus the name 74:synthetic polymers 2467: 2466: 2206: 2205: 2182:Molecular docking 2011:Mass spectrometry 2006:Fiber diffraction 1999:Medium resolution 1898:Library resources 1847:10.1002/app.40111 1711:978-93-80473-15-4 1679:10.1021/ma902377g 1634:10.1021/ma030586k 1620:(11): 4304–4312. 1415:10.1038/1831657a0 1337:10.1042/bj0620665 1143:10.1021/es000076v 1129:(23): 4867–4872. 1038:978-0-470-68218-0 960:10.1021/ef700410e 779:calibration curve 773:of a synthesized 765:Polymer synthesis 726:of the molecular 648: 647: 382:Theory and method 118: 117: 16:(Redirected from 2512: 2457: 2456: 2404:Retention factor 2233: 2226: 2219: 2210: 2083:Light scattering 1949: 1942: 1935: 1926: 1885: 1884: 1858: 1852: 1851: 1849: 1825: 1819: 1818: 1794: 1788: 1787: 1777: 1751: 1742: 1741: 1739: 1737: 1722: 1716: 1715: 1697: 1691: 1690: 1665:(3): 1651–1659. 1652: 1646: 1645: 1607: 1601: 1600: 1562: 1556: 1555: 1529: 1523: 1522: 1492: 1486: 1485: 1467: 1441: 1435: 1434: 1401:(4676): 1657–9. 1390: 1384: 1383: 1381: 1379: 1365: 1359: 1358: 1348: 1316: 1310: 1309: 1299: 1275: 1269: 1268: 1250: 1241: 1240: 1216: 1206: 1200: 1199: 1167: 1158: 1147: 1146: 1116: 1110: 1109: 1099: 1081: 1049: 1043: 1042: 1024: 1018: 1017: 1007: 997: 973: 964: 963: 954:(6): 3484–3489. 941: 935: 934: 908: 709:refractive index 464: 328: 326: 325: 320: 315: 314: 302: 301: 289: 281: 280: 268: 267: 252: 251: 157:aqueous solution 123:, also known as 92:Other techniques 40: 33: 21: 2520: 2519: 2515: 2514: 2513: 2511: 2510: 2509: 2470: 2469: 2468: 2463: 2445: 2413: 2377: 2351: 2260: 2242: 2237: 2207: 2202: 2201: 2196: 2186: 2165: 2149: 2123: 2092: 2061: 2020: 1994: 1963:High resolution 1958: 1953: 1923: 1922: 1921: 1906: 1905: 1901: 1894: 1889: 1888: 1873: 1860: 1859: 1855: 1827: 1826: 1822: 1796: 1795: 1791: 1753: 1752: 1745: 1735: 1733: 1724: 1723: 1719: 1712: 1699: 1698: 1694: 1654: 1653: 1649: 1609: 1608: 1604: 1564: 1563: 1559: 1544: 1531: 1530: 1526: 1494: 1493: 1489: 1443: 1442: 1438: 1392: 1391: 1387: 1377: 1375: 1367: 1366: 1362: 1318: 1317: 1313: 1277: 1276: 1272: 1265: 1252: 1251: 1244: 1229: 1208: 1207: 1203: 1180: 1165: 1160: 1159: 1150: 1118: 1117: 1113: 1051: 1050: 1046: 1039: 1026: 1025: 1021: 975: 974: 967: 943: 942: 938: 923: 910: 909: 902: 897: 875: 856: 852: 839: 818: 800: 767: 745: 740: 681: 653: 553:Sephadex G-200 542:Sephadex G-150 531:Sephadex G-100 472:And Trade name 384: 348: 340: 336: 332: 306: 293: 272: 259: 240: 235: 234: 232: 222: 201: 151:and industrial 129:chromatographic 111: 107: 103: 76: 72: 43: 28: 23: 22: 15: 12: 11: 5: 2518: 2516: 2508: 2507: 2502: 2497: 2492: 2487: 2482: 2480:Chromatography 2472: 2471: 2465: 2464: 2462: 2461: 2450: 2447: 2446: 2444: 2443: 2436: 2429: 2421: 2419: 2415: 2414: 2412: 2411: 2406: 2401: 2396: 2391: 2385: 2383: 2379: 2378: 2376: 2375: 2370: 2365: 2359: 2357: 2353: 2352: 2350: 2349: 2344: 2339: 2334: 2329: 2324: 2319: 2314: 2309: 2304: 2299: 2294: 2289: 2284: 2279: 2274: 2268: 2266: 2262: 2261: 2259: 2258: 2253: 2247: 2244: 2243: 2240:Chromatography 2238: 2236: 2235: 2228: 2221: 2213: 2204: 2203: 2197: 2192: 2191: 2188: 2187: 2185: 2184: 2179: 2173: 2171: 2167: 2166: 2164: 2163: 2157: 2155: 2151: 2150: 2148: 2147: 2142: 2137: 2131: 2129: 2125: 2124: 2122: 2121: 2116: 2111: 2106: 2100: 2098: 2094: 2093: 2091: 2090: 2085: 2080: 2075: 2069: 2067: 2063: 2062: 2060: 2059: 2054: 2049: 2044: 2039: 2034: 2028: 2026: 2022: 2021: 2019: 2018: 2013: 2008: 2002: 2000: 1996: 1995: 1993: 1992: 1987: 1982: 1977: 1972: 1966: 1964: 1960: 1959: 1954: 1952: 1951: 1944: 1937: 1929: 1920: 1919: 1914: 1908: 1907: 1896: 1895: 1893: 1892:External links 1890: 1887: 1886: 1871: 1853: 1820: 1789: 1743: 1717: 1710: 1692: 1658:Macromolecules 1647: 1613:Macromolecules 1602: 1575:(9): 753–759. 1557: 1542: 1524: 1505:(2): 835–843. 1487: 1451:Nature Methods 1436: 1385: 1360: 1311: 1270: 1263: 1242: 1228:978-0471972617 1227: 1201: 1178: 1148: 1111: 1064:(13): 7064–7. 1044: 1037: 1019: 965: 936: 921: 899: 898: 896: 893: 892: 891: 886: 881: 874: 871: 854: 850: 837: 817: 814: 799: 796: 771:polydispersity 766: 763: 744: 741: 739: 736: 680: 677: 652: 649: 646: 645: 642: 639: 635: 634: 631: 628: 624: 623: 620: 617: 613: 612: 609: 608:Bio-gel P-300 606: 602: 601: 598: 597:Bio-gel P-150 595: 591: 590: 587: 584: 580: 579: 576: 573: 569: 568: 565: 562: 558: 557: 554: 551: 547: 546: 543: 540: 536: 535: 532: 529: 525: 524: 521: 520:Sephadex G-75 518: 514: 513: 510: 509:Sephadex G-50 507: 503: 502: 499: 498:Sephadex G-25 496: 492: 491: 488: 487:Sephadex G-10 485: 481: 480: 474: 468: 383: 380: 347: 344: 338: 334: 330: 318: 313: 309: 305: 300: 296: 292: 288: 284: 279: 275: 271: 266: 262: 258: 255: 250: 247: 243: 230: 221: 218: 200: 197: 181:macromolecules 177:polyacrylamide 116: 115: 98: 94: 93: 89: 88: 85: 81: 80: 70:macromolecules 67: 63: 62: 60:Chromatography 57: 56:Classification 53: 52: 49: 45: 44: 41: 26: 24: 14: 13: 10: 9: 6: 4: 3: 2: 2517: 2506: 2503: 2501: 2498: 2496: 2493: 2491: 2488: 2486: 2483: 2481: 2478: 2477: 2475: 2460: 2452: 2451: 2448: 2442: 2441: 2437: 2435: 2434: 2430: 2428: 2427: 2423: 2422: 2420: 2416: 2410: 2407: 2405: 2402: 2400: 2397: 2395: 2392: 2390: 2387: 2386: 2384: 2380: 2374: 2371: 2369: 2366: 2364: 2361: 2360: 2358: 2354: 2348: 2345: 2343: 2340: 2338: 2335: 2333: 2330: 2328: 2325: 2323: 2320: 2318: 2315: 2313: 2310: 2308: 2305: 2303: 2300: 2298: 2295: 2293: 2290: 2288: 2285: 2283: 2280: 2278: 2275: 2273: 2270: 2269: 2267: 2263: 2257: 2254: 2252: 2249: 2248: 2245: 2241: 2234: 2229: 2227: 2222: 2220: 2215: 2214: 2211: 2200: 2195: 2189: 2183: 2180: 2178: 2175: 2174: 2172: 2170:Computational 2168: 2162: 2159: 2158: 2156: 2154:Thermodynamic 2152: 2146: 2143: 2141: 2138: 2136: 2133: 2132: 2130: 2126: 2120: 2117: 2115: 2112: 2110: 2107: 2105: 2102: 2101: 2099: 2095: 2089: 2086: 2084: 2081: 2079: 2076: 2074: 2071: 2070: 2068: 2064: 2058: 2055: 2053: 2050: 2048: 2045: 2043: 2040: 2038: 2035: 2033: 2030: 2029: 2027: 2025:Spectroscopic 2023: 2017: 2014: 2012: 2009: 2007: 2004: 2003: 2001: 1997: 1991: 1988: 1986: 1983: 1981: 1978: 1976: 1973: 1971: 1968: 1967: 1965: 1961: 1957: 1950: 1945: 1943: 1938: 1936: 1931: 1930: 1927: 1918: 1915: 1913: 1910: 1909: 1904: 1899: 1891: 1882: 1878: 1874: 1872:9781904455462 1868: 1864: 1857: 1854: 1848: 1843: 1839: 1835: 1831: 1824: 1821: 1816: 1812: 1808: 1804: 1800: 1793: 1790: 1785: 1781: 1776: 1775:10.3791/59615 1771: 1767: 1763: 1762: 1757: 1750: 1748: 1744: 1731: 1727: 1721: 1718: 1713: 1707: 1703: 1696: 1693: 1688: 1684: 1680: 1676: 1672: 1668: 1664: 1660: 1659: 1651: 1648: 1643: 1639: 1635: 1631: 1627: 1623: 1619: 1615: 1614: 1606: 1603: 1598: 1594: 1590: 1586: 1582: 1578: 1574: 1570: 1569: 1568:J Polym Sci B 1561: 1558: 1553: 1549: 1545: 1543:9780470442876 1539: 1535: 1528: 1525: 1520: 1516: 1512: 1508: 1504: 1500: 1499: 1498:J Polym Sci A 1491: 1488: 1483: 1479: 1475: 1471: 1466: 1461: 1457: 1453: 1452: 1447: 1440: 1437: 1432: 1428: 1424: 1420: 1416: 1412: 1408: 1404: 1400: 1396: 1389: 1386: 1374: 1370: 1364: 1361: 1356: 1352: 1347: 1342: 1338: 1334: 1331:(4): 665–74. 1330: 1326: 1322: 1315: 1312: 1307: 1303: 1298: 1293: 1289: 1285: 1281: 1274: 1271: 1266: 1264:9780470087664 1260: 1256: 1249: 1247: 1243: 1238: 1234: 1230: 1224: 1220: 1215: 1214: 1205: 1202: 1197: 1193: 1189: 1185: 1181: 1179:9780495012016 1175: 1171: 1164: 1157: 1155: 1153: 1149: 1144: 1140: 1136: 1132: 1128: 1124: 1123: 1115: 1112: 1107: 1103: 1098: 1093: 1089: 1085: 1080: 1075: 1071: 1067: 1063: 1059: 1055: 1048: 1045: 1040: 1034: 1030: 1023: 1020: 1015: 1011: 1006: 1001: 996: 991: 987: 983: 979: 972: 970: 966: 961: 957: 953: 949: 948: 940: 937: 932: 928: 924: 922:9781133106296 918: 914: 907: 905: 901: 894: 890: 887: 885: 882: 880: 877: 876: 872: 870: 866: 862: 858: 848: 842: 833: 831: 827: 823: 815: 813: 810: 804: 797: 795: 793: 788: 784: 780: 776: 772: 764: 762: 759: 755: 751: 742: 737: 735: 733: 729: 725: 720: 718: 714: 710: 706: 697: 689: 685: 678: 676: 672: 670: 662: 657: 650: 643: 641:Sepharose 6B 640: 637: 636: 632: 630:Sepharose 4B 629: 626: 625: 621: 619:Sepharose 2B 618: 615: 614: 610: 607: 604: 603: 599: 596: 593: 592: 588: 586:Bio-gel P-60 585: 582: 581: 577: 574: 571: 570: 566: 563: 560: 559: 555: 552: 549: 548: 544: 541: 538: 537: 533: 530: 527: 526: 522: 519: 516: 515: 511: 508: 505: 504: 500: 497: 494: 493: 489: 486: 483: 482: 479: 475: 473: 469: 466: 465: 460: 457: 455: 454:column volume 451: 447: 442: 434: 430: 426: 424: 419: 414: 412: 406: 404: 396: 392: 388: 381: 379: 377: 373: 369: 365: 360: 358: 354: 353:Jerker Porath 345: 343: 311: 307: 303: 298: 294: 286: 277: 273: 269: 264: 260: 253: 248: 245: 241: 226: 219: 217: 213: 211: 206: 205:fractionation 198: 196: 192: 190: 186: 182: 178: 174: 170: 166: 162: 158: 154: 150: 146: 142: 138: 134: 130: 126: 122: 114: 110: 106: 102: 99: 95: 90: 86: 84:Manufacturers 82: 79: 75: 71: 68: 64: 61: 58: 54: 50: 46: 39: 34: 19: 2438: 2431: 2424: 2336: 2052:Fluorescence 1902: 1862: 1856: 1840:(7): 40111. 1837: 1833: 1823: 1806: 1802: 1792: 1765: 1759: 1734:. Retrieved 1729: 1720: 1701: 1695: 1662: 1656: 1650: 1617: 1611: 1605: 1572: 1566: 1560: 1533: 1527: 1502: 1496: 1490: 1455: 1449: 1439: 1398: 1394: 1388: 1376:. Retrieved 1372: 1363: 1328: 1324: 1314: 1290:(4): xxxiv. 1287: 1283: 1273: 1254: 1212: 1204: 1169: 1126: 1120: 1114: 1061: 1057: 1047: 1028: 1022: 985: 981: 951: 945: 939: 913:Biochemistry 912: 867: 863: 859: 843: 834: 819: 805: 801: 768: 746: 738:Applications 721: 702: 682: 673: 665: 660: 575:Bio-gel P-6 564:Bio-gel P-2 477: 471: 458: 453: 449: 445: 443: 439: 427: 415: 407: 400: 361: 349: 227: 223: 214: 202: 199:Applications 193: 160: 124: 120: 119: 78:biomolecules 2505:Polyolefins 1809:(1): 1–40. 732:void volume 622:2000-25000 450:void volume 372:polystyrene 2474:Categories 2265:Techniques 2047:Absorbance 1458:(5): 410. 1188:2006926952 931:1066452448 895:References 879:PEGylation 787:viscometry 357:Per Flodin 220:Advantages 1881:430080586 1687:0024-9297 1642:0024-9297 1597:1542-6254 1552:587401945 1519:1542-6246 1474:1548-7105 1237:635171657 988:: 44–48. 798:Drawbacks 724:logarithm 644:10-20000 633:300-3000 470:Material 403:adsorbent 346:Discovery 304:− 270:− 145:molecules 2500:Polymers 2459:Category 2251:software 2128:Chemical 1784:31282880 1482:37935968 1431:32287460 1423:13666849 1355:13315231 1306:13249976 1196:77224390 1106:10852951 1014:18949074 873:See also 809:friction 679:Analysis 600:1.5-150 567:0.1-1.8 153:polymers 149:proteins 133:solution 66:Analytes 2256:history 1768:(148). 1730:pslc.ws 1667:Bibcode 1622:Bibcode 1577:Bibcode 1403:Bibcode 1346:1215979 1297:1216175 1131:Bibcode 1066:Bibcode 1005:2570550 775:polymer 611:16-400 556:5-8000 512:1.5-30 467:Sr. No 397:machine 391:Agarose 173:agarose 169:dextran 127:, is a 97:Related 48:Acronym 2382:Theory 1900:about 1879:  1869:  1782:  1708:  1685:  1640:  1595:  1550:  1540:  1517:  1480:  1472:  1429:  1421:  1395:Nature 1353:  1343:  1304:  1294:  1261:  1235:  1225:  1221:–103. 1194:  1186:  1176:  1104:  1094:  1088:122767 1086:  1035:  1012:  1002:  929:  919:  545:5-300 534:4-150 490:0-0.7 478:(kDa) 448:. The 446:eluate 1736:3 Jan 1478:S2CID 1427:S2CID 1378:3 Jan 1166:(PDF) 1097:16499 1084:JSTOR 950:. 6. 713:LALLS 589:3-60 523:3-70 411:elute 175:, or 137:shape 2016:SAXS 1877:OCLC 1867:ISBN 1780:PMID 1766:2019 1738:2019 1706:ISBN 1683:ISSN 1638:ISSN 1593:ISSN 1548:OCLC 1538:ISBN 1515:ISSN 1470:ISSN 1419:PMID 1380:2019 1351:PMID 1302:PMID 1259:ISBN 1233:OCLC 1223:ISBN 1192:OCLC 1184:LCCN 1174:ISBN 1102:PMID 1033:ISBN 1010:PMID 927:OCLC 917:ISBN 578:1-6 501:1-5 395:FPLC 355:and 141:size 2119:NMR 2088:NMR 2032:NMR 1990:EPR 1980:NMR 1842:doi 1838:131 1811:doi 1807:272 1770:doi 1675:doi 1630:doi 1585:doi 1507:doi 1460:doi 1411:doi 1399:183 1341:PMC 1333:doi 1292:PMC 1219:101 1139:doi 1092:PMC 1074:doi 1000:PMC 990:doi 956:doi 783:THF 638:15 627:14 616:13 605:12 594:11 583:10 51:SEC 2476:: 1875:. 1836:. 1832:. 1805:. 1801:. 1778:. 1764:. 1758:. 1746:^ 1728:. 1681:. 1673:. 1663:43 1661:. 1636:. 1628:. 1618:37 1616:. 1591:. 1583:. 1571:. 1546:. 1513:. 1501:. 1476:. 1468:. 1454:. 1448:. 1425:. 1417:. 1409:. 1397:. 1371:. 1349:. 1339:. 1329:62 1327:. 1323:. 1300:. 1288:60 1286:. 1282:. 1245:^ 1231:. 1190:. 1182:. 1168:. 1151:^ 1137:. 1127:34 1125:. 1100:. 1090:. 1082:. 1072:. 1062:97 1060:. 1056:. 1008:. 998:. 984:. 980:. 968:^ 952:21 925:. 903:^ 836:(D 715:, 572:9 561:8 550:7 539:6 528:5 517:4 506:3 495:2 484:1 456:. 231:av 171:, 2232:e 2225:t 2218:v 1948:e 1941:t 1934:v 1883:. 1850:. 1844:: 1817:. 1813:: 1786:. 1772:: 1740:. 1714:. 1689:. 1677:: 1669:: 1644:. 1632:: 1624:: 1599:. 1587:: 1579:: 1573:5 1554:. 1521:. 1509:: 1503:2 1484:. 1462:: 1456:3 1433:. 1413:: 1405:: 1382:. 1357:. 1335:: 1308:. 1267:. 1239:. 1198:. 1145:. 1141:: 1133:: 1108:. 1076:: 1068:: 1041:. 1016:. 992:: 986:2 962:. 958:: 933:. 855:g 851:g 838:H 339:o 335:t 331:e 317:) 312:o 308:V 299:t 295:V 291:( 287:/ 283:) 278:o 274:V 265:e 261:V 257:( 254:= 249:v 246:a 242:K 20:)

Index

Size exclusion chromatography

Chromatography
macromolecules
synthetic polymers
biomolecules
High-performance liquid chromatography
Aqueous normal-phase chromatography
Ion exchange chromatography
Micellar liquid chromatography
chromatographic
solution
shape
size
molecules
proteins
polymers
aqueous solution
gel permeation chromatography
dextran
agarose
polyacrylamide
macromolecules
polymer characterization
molar mass distribution
fractionation
gel electrophoresis
Jerker Porath
Per Flodin
Dow Chemical Company

Text is available under the Creative Commons Attribution-ShareAlike License. Additional terms may apply.

↑