Knowledge (XXG)

Glucose meter

Source đź“ť

596:(NHS) rather than patients pay for medications including test strips, a 2015 study on the comparative cost-effectiveness of all options for the self-monitoring of blood glucose funded by the NHS uncovered considerable variation in the price charged, which could not be explained by the availability of advanced meter features. It estimated that a total of ÂŁ12m was invested in providing 42 million self-monitoring blood glucose tests with systems that failed to meet acceptable accuracy standards, and efficiency savings of ÂŁ23.2m per annum were achievable if the NHS were to disinvest from technologies providing less functionality than available alternatives, but at a much higher price. 1036:
28 mmol/L), and when a meter reading of 50 or 70 (2.8 or 3.9 mmol/L) is accompanied by their usual hypoglycemic symptoms, there is little uncertainty about the reading representing a "true positive" and little harm done if it is a "false positive." However, the incidence of hypoglycemia unawareness, hypoglycemia-associated autonomic failure (HAAF) and faulty counterregulatory response to hypoglycemia make the need for greater reliability at low levels particularly urgent in patients with type 1 diabetes mellitus, while this is seldom an issue in the more common form of the disease, type 2 diabetes mellitus.
509:: The glucose value in mg/dL or mmol/L (1 mmol/L = 18.0 mg/dL) is displayed on a digital display. Different countries use different measurement units: for example mg/dL are used in the US, France, Japan, Iran, Israel, and India; mmol/L are used in Australia, Canada, China, and the UK. In Germany both units are used. Many meters can display either unit of measure. Instances have been published in which a patient has interpreted a reading in mmol/L as a very low reading in mg/dL or vice versa. Usually mmol/L readings have a decimal point and mg/dL readings do not. 252: 385: 377: 809: 697: 735:. The device was designed to be worn on the wrist and used electric fields to draw out body fluid for testing. The device did not replace conventional blood glucose monitoring. One limitation was that the GlucoWatch was not able to cope with perspiration at the measurement site. Sweat must be allowed to dry before measurement can resume. Due to this limitation and others, the product is no longer on the market. 481:: Since test strips may vary from batch to batch, some models require a code to be provided, either by the user or on a plug-in chip supplied with each batch of test strips, to calibrate the meter to the strips of the batch. An incorrect code can cause errors of up to 4 mmol/L (72 mg/dL), with possibly serious consequences, including risk of hypoglycemia. Some test media contain the code information in the strip. 894: 398: 36: 658:(ISO). According to ISO 15197 Blood glucose meters must provide results that are within ±15% of a laboratory standard for concentrations above 100 mg/dL or within ±15 mg/dL for concentrations below 100 mg/dL at least 95% of the time. However, a variety of factors can affect the accuracy of a test. Factors affecting accuracy of various meters include calibration of meter, 532:
plasma equivalent is calculated from the whole blood glucose reading using an equation built into the glucose meter. This allows patients to easily compare their glucose measurements in a lab test and at home. It is important for patients and their health care providers to know whether the meter gives its results as "whole blood equivalent" or "plasma equivalent." One model measures
960: 142: 340:
manufacturer cited studies that show the product is just as effective despite not giving an answer to one decimal place, something they argue is unnecessary for control of blood sugar. This debate also happened in Germany where "Glucoflex-R" was an established strip for type 2 diabetes. As meter accuracy and insurance coverage improved, they lost popularity.
627: 752:
data directly to a compatible cellular telephone. Currently, Apple's iPhone and Android devices can be used as a receiver. Aside from a two-hour calibration period, monitoring is logged at five-minute intervals for up to 1 week. The user can set the high and low glucose alarms. The third CGMS available is the FreeStyle Navigator from Abbott Laboratories.
248:(GOx) on an oxygen electrode. Thus, the readout was the amount of oxygen consumed by GOx during the enzymatic reaction with the substrate glucose. This publication became one of the most often cited papers in life sciences. Due to this work he is considered the “father of biosensors,” especially with respect to the glucose sensing for diabetes patients. 869:
throwing a ball and using the speed at which it is travelling at a point in time to estimate how hard it was thrown. The coulometric method can allow for variable test times, whereas the test time on a meter using the amperometric method is always fixed. Both methods give an estimation of the concentration of glucose in the initial blood sample.
1234:; Ginsberg, Barry H.; Alva, Shridhara; Appel, Matthias; Bess, Stephan; Chen, Kong Y.; Freckmann, Guido; Harris, Dennis R.; Hartwig, Matthias; Hinzmann, Rolf; Kerr, David; Krouwer, Jan; Morrow, Linda; Nichols, James; PfĂĽtzner, Andreas; Pleus, Stefan; Rice, Mark; Sacks, David B.; Schlueter, Kevin; Vesper, Hubert W.; Klonoff, David C. (May 2020). 861:
regeneration of the oxidised form of the mediator for shuttling of electrons from enzyme to active site. Osmium-based polypyridyl redox complexes and polymers are attractive candidates as mediators due to their stability in oxidised and reduced forms, tunable redox potential, ease of co-immobilisation and ability to operate at low potentials.
343:"Glucoflex-R" is Australia manufacturer National Diagnostic Products alternative to the BM test strip. It has versions that can be used either in a meter or read visually. It is also marketed under the brand name Betachek. On May 1, 2009, the UK distributor Ambe Medical Group reduced the price of their "Glucoflex-R" test strip to the 1032:. An imprecision of ±15% is less of a problem for high glucose levels than low. There is little difference in the management of a glucose of 200 mg/dL compared with 260 (i.e., a "true" glucose of 230±15%), but a ±15% error margin at a low glucose concentration brings greater ambiguity with regards to glucose management. 244:, on 15 April 1956, at a meeting of the American Society for Artificial Organs during the annual meetings of the Federated Societies for Experimental Biology. In 1962, Clark and Ann Lyons from the Cincinnati Children's Hospital developed the first glucose enzyme electrode. This biosensor was based on a thin layer of 497:: Smaller drop volumes have enabled "alternate site testing" – pricking the forearms or other less sensitive areas instead of the fingertips. Manufacturers recommend that this type of testing should only be used when blood glucose levels are stable, such as before meals, when fasting, or just before going to sleep. 872:
The same principle is used in test strips that have been commercialized for the detection of diabetic ketoacidosis (DKA). These test strips use a beta-hydroxybutyrate-dehydrogenase enzyme instead of a glucose oxidizing enzyme and have been used to detect and help treat some of the complications that
617:
The search for a successful technique began about 1975 and has continued to the present without a clinically or commercially viable product. As of 1999, only one such product had ever been approved for sale by the FDA, based on a technique for electrically pulling glucose through intact skin, and it
1039:
In contrast, people who do not have diabetes may periodically have hypoglycemic symptoms but may also have a much higher rate of false positives to true, and a meter is not accurate enough to base a diagnosis of hypoglycemia upon. A meter can occasionally be useful in the monitoring of severe types
738:
The market introduction of noninvasive blood glucose measurement by spectroscopic measurement methods, in the field of near-infrared (NIR), by extracorporal measuring devices, has not been successful because the devices measure tissue sugar in body tissues and not the blood sugar in blood fluid. To
868:
method is a technique where the total amount of charge generated by the glucose oxidation reaction is measured over a period of time. The amperometric method is used by some meters and measures the electric current generated at a specific point in time by the glucose reaction. This is analogous to
851:
Most glucometers today use an electrochemical method. Test strips contain a capillary that sucks up a reproducible amount of blood. The glucose in the blood reacts with an enzyme electrode containing glucose oxidase (or dehydrogenase). The enzyme is reoxidized with an excess of a mediator reagent,
751:
probe with a small transmitter. The receiver is about the size of a cell phone and can operate up to twenty feet from the transmitter. The Dexcom G4 transmits via radio frequency and requires a dedicated receiver. The G5 version utilizes Bluetooth low energy for data transmission, and can transmit
513:
Countries that use mmol/L include Australia, Canada, China, Croatia, Czech Republic, Denmark, Finland, Hong Kong, Hungary, Iceland, Ireland, Jamaica, Kazakhstan, Latvia, Lithuania, Malaysia, Malta, Netherlands, New Zealand, Norway, Russia, Slovakia, Slovenia, South Africa, Sweden, Switzerland, and
591:
0.35 to $ 1.00. Manufacturers often provide meters at no cost to encourage use of the profitable test strips. Type 1 diabetics may test as often as 4 to 10 times a day due to the dynamics of insulin adjustment, whereas type 2 typically test less frequently, especially when insulin is not part of
339:
Test strips that changed color and could be read visually, without a meter, have been widely used since the 1980s. They had the added advantage that they could be cut longitudinally to save money. Critics argued that test strips read by eye are not as accurate or convenient as meter testing. The
787:
Recent advances in cellular data communications technology have enabled the development of glucose meters that directly integrate cellular data transmission capability, enabling the user to both transmit glucose data to the medical caregiver and receive direct guidance from the caregiver on the
746:
RTS with a sub-cutaneous probe attached to a small transmitter (roughly the size of a quarter) that sends interstitial glucose levels to a small pager sized receiver every five minutes. The Dexcom System is another system, available in two different generations in the US, the G4 and the G5. (1Q
531:
is normal. This is important because home blood glucose meters measure the glucose in whole blood while most lab tests measure the glucose in plasma. Currently, there are many meters on the market that give results as "plasma equivalent," even though they are measuring whole blood glucose. The
517:
Countries that use mg/dL include Algeria, Argentina, Austria, Bangladesh, Belgium, Brazil, Chile, Columbia, Cyprus, Ecuador, Egypt, France, Georgia, Germany, Greece, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Korea, Lebanon, Mexico, Peru, Poland, Portugal, South Korea, Spain, Syria,
1035:
The imprecision is compounded by the relative likelihoods of false positives and negatives in populations with diabetes and those without. People with type 1 diabetes usually have a wider range of glucose levels, and glucose peaks above normal, often ranging from 40 to 500 mg/dL (2.2 to
860:
derivative or osmium bipyridyl complex. The mediator in turn is reoxidized by reaction at the electrode, which generates an electric current. In order for the mediator to operate over long timeframes, it needs to be stable in both oxidised and reduced states. This is to allow for continuous
641:
Continuous glucose monitor systems can consist of a disposable sensor placed under the skin, a transmitter connected to the sensor and a reader that receives and displays the measurements. The sensor can be used for several days before it needs to be replaced. The devices provide real-time
491:: The size of the drop of blood needed by different models varies from 0.3 to 1 ÎĽl. Older models required larger blood samples, usually defined as a "hanging drop" from the fingertip. Smaller volume requirements reduce the frequency of pricks that do not produce enough blood. 763:
controller, as well as an effort to integrate the glucose meter and a cell phone. Testing strips are proprietary and available only through the manufacturer (no insurance availability). These "Glugophones" are currently offered in three forms: as a dongle for the
471:, which reacts with glucose, and other components. A drop of blood is absorbed by a spot for each measurement. Some models use single-use plastic test strips with a spot; other models use discs, drums, or cartridges with multiple spots to make several readings. 330:
more slowly than for type 1, and a large proportion of people with type 2 diabetes have never been instructed in home glucose testing. This has mainly come about because health authorities are reluctant to bear the cost of the test strips and lancets.
347:, by approximately 50%. This was expected to allow the NHS to save money on strips and perhaps loosen the restrictions on supply a little. Another low cost visually read strip is soon to be available on prescription according to sources at the NHS. 799:
that monitor glucose levels and alert users when glucose levels cross certain thresholds. Apple has patented methods for determining blood sugar levels by absorption spectroscopy, as well as by analyzing exhaled air in its electronic devices.
1548:
Leigh, Simon; Idris, Iskandar; Collins, Brendan; Granby, Paul; Noble, Max; Parker, Mark (Nov 2015). "Promoting health and reducing costs: a role for reform of self-monitoring of blood glucose provision within the National Health Service".
310:
argued that the superior accuracy of a laboratory glucose measurement outweighed the advantage of immediate availability and made meter glucose measurements unacceptable for inpatient diabetes management. Patients with diabetes and their
788:
screen of the glucose meter. The first such device, from Telcare, Inc., was exhibited at the 2010 CTIA International Wireless Expo, where it won an E-Tech award. This device then underwent clinical testing in the US and internationally.
270:
of type 1 diabetes in the late 1970s, and the first meters were marketed for home use around 1981. The two models initially dominant in North America in the 1980s were the Glucometer, introduced in November 1981, whose
847:
formed in the oxidation reaction. The disadvantage of this method was that the test strip had to be developed after a precise interval (the blood had to be washed away), and the meter needed to be calibrated frequently.
216:
has been achieving closer-to-normal levels of glucose in the blood for as much of the time as possible, guided by HBGM several times a day. The benefits include a reduction in the occurrence rate and severity of
1713:"Analytical Performance Requirements for Systems for Self-Monitoring of Blood Glucose With Focus on System Accuracy: Relevant Differences Among ISO 15197:2003, ISO 15197:2013, and Current FDA Recommendations" 1592: 678:
has been a common way of analyzing and displaying accuracy of readings related to management consequences. More recently an improved version of the Clarke Error Grid has come into use: It is known as the
630:
Continuous glucose monitor. The sensor and transmitter are fixed to the upper arm. The reader shows days to replacement of sensor, current blood glucose level and a diagram of the latest blood glucose
360:
Special glucose meters for multi-patient hospital use are now used. These provide more elaborate quality control records. Their data handling capabilities are designed to transfer glucose results into
146:
Four generations of blood glucose meter, c. 1993–2005. Sample sizes vary from 30 to 0.3 μl. Test times vary from 5 seconds to 2 minutes (modern meters typically provide results in 5 seconds).
2009:
Ghoshdastider U, Wu R, Trzaskowski B, Mlynarczyk K, Miszta P, Gurusaran M, Viswanathan S, Renugopalakrishnan V, Filipek S (2015). "Nano-Encapsulation of Glucose Oxidase Dimer by Graphene".
683:. Older blood glucose meters often need to be "coded" with the lot of test strips used, otherwise, the accuracy of the blood glucose meter may be compromised due to lack of calibration. 2041: 1098:
Advances in Electrochemical Sciences and Engineering : Bioelectrochemistry : Fundamentals, Applications and Recent Developments. Somerset, NJ, US: John Wiley & Sons, 2013.
554:
by a clock set by the user, and many can display an average of recent readings. Stored data will only reflect trends accurately if the clock is set to approximately the right time.
880:
Blood alcohol sensors using the same approach, but with alcohol dehydrogenase enzymes, have been tried and patented but have not yet been successfully commercially developed.
642:
measurements, and reduce the need for fingerprick testing of glucose levels. A drawback is that the meters are not as accurate because they read the glucose levels in the
587:
The cost of home blood glucose monitoring can be substantial due to the cost of the test strips. In 2006, the US cost to consumers of each glucose strip ranged from about
674:, and aging of test strips. Models vary in their susceptibility to these factors and in their ability to prevent or warn of inaccurate results with error messages. The 603:
test strips for some meters were found in the United States, producing erratic test results that do not meet the legitimate manufacturer's performance specifications.
1496: 316: 655: 201:, is placed on a disposable test strip that the meter reads and uses to calculate the blood glucose level. The meter then displays the level in units of mg/dL or 2056: 1915: 1110:
Lipkowski, J., Kolb, D. M., & Alkire, R. C. (2011). Bioelectrochemistry : Fundamentals, Applications and Recent Developments. Weinheim: Wiley-VCH.
1865: 527:: Glucose levels in plasma (one of the components of blood) are higher than glucose measurements in whole blood; the difference is about 11% when the 1662: 1285: 1478: 1801:"CTIA International Wireless 2010 Finds Telcare Inc. Measures Up to Emerging Technology Award for Cellular-enabled Blood Glucose Meter Solution" 1593:"FDA Warns About Fake Diabetic Blood Test Strips Sold For Use With Various Models Of LifeScan, Inc.'s One Touch Brand Blood Glucose Monitors" 1517: 864:
The total charge passing through the electrode is proportional to the amount of glucose in the blood that has reacted with the enzyme. The
1181:"Fundamental Importance of Reference Glucose Analyzer Accuracy for Evaluating the Performance of Blood Glucose Monitoring Systems (BGMSs)" 832:(GDH). This has the advantage of sensitivity over glucose oxidase but is more susceptible to interfering reactions with other substances. 739:
determine blood glucose, the measuring beam of infrared light, for example, has to penetrate the tissue for measurement of blood glucose.
263:
by Anton H. Clemens. It was used in American hospitals in the 1970s. A moving needle indicated the blood glucose after about a minute.
2051: 2046: 1179:
Bailey, Timothy S.; Klaff, Leslie J.; Wallace, Jane F.; Greene, Carmine; Pardo, Scott; Harrison, Bern; Simmons, David A. (July 2016).
1003: 941: 437: 119: 1154: 970: 829: 503:: The time it takes for a reading to be displayed may range from 3 to 60 seconds from application of blood for different models. 662:, pressure use to wipe off strip (if applicable), size and quality of blood sample, high levels of certain substances (such as 654:
Accuracy of glucose meters is a common topic of clinical concern. Blood glucose meters must meet accuracy standards set by the
919: 915: 419: 415: 181:. It can also be a strip of glucose paper dipped into a substance and measured to the glucose chart. It is a key element of 57: 53: 100: 72: 2061: 1610: 839:
reaction that is still used nowadays in glucose test strips for urine. Besides glucose oxidase, the test kit contains a
569: 319:
was widely accepted as the de facto standard for reference measurements and system calibration by most manufacturers of
646:
which lags behind the levels in the blood. Continuous blood glucose monitoring systems are also relatively expensive.
612: 1919: 985: 904: 79: 1041: 636: 573: 981: 923: 908: 408: 46: 361: 251: 213: 209: 186: 86: 384: 376: 1286:"Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus™ and YSI 2900D Biochemistry Analyzers" 295:, a health care professional or a patient may refer to "taking a BM": "Mrs X's BM is 5", etc. BM stands for 1821:
NM Farandos; AK Yetisen; MJ Monteiro; CR Lowe; SH Yun (2014). "Contact Lens Sensors in Ocular Diagnostics".
593: 344: 306:
resisted adoption of meter glucose measurements for inpatient diabetes care for over a decade. Managers of
742:
There are currently three CGMS (continuous glucose monitoring system) available. The first is Medtronic's
260: 68: 1449: 280: 1780: 808: 458: 288: 226: 618:
was withdrawn after a short time owing to poor performance and occasional damage to the skin of users.
518:
Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United States, Uruguay, Venezuela, and Yemen.
1624: 1482: 731:
One noninvasive glucose meter has been approved by the U.S. FDA: The GlucoWatch G2 Biographer made by
1941:
Ruff, Adrian (2017). "Redox polymers in bioelectrochemistry: Common playgrounds and novel concepts".
680: 533: 1425: 659: 296: 267: 1524: 1016:
Although the apparent value of immediate measurement of blood glucose might seem to be higher for
1991: 1846: 1574: 1044:) to ensure that the average glucose when fasting remains above 70 mg/dL (3.9 mmol/L). 643: 365: 307: 812:
Two used Accu-Chek test strips. The lower one has had the cover peeled off to show the circuit.
1838: 1742: 1644: 1566: 1386: 1335: 1267: 1212: 844: 675: 450:
There are several key characteristics of glucose meters which may differ from model to model:
218: 190: 755:
There is currently an effort to develop an integrated treatment system with a glucose meter,
2018: 1981: 1950: 1830: 1732: 1724: 1636: 1558: 1376: 1366: 1325: 1315: 1257: 1247: 1202: 1192: 743: 1893: 1405:"Diabetes UK, UK Diabetes Resource, Diabetes Symptoms, Diabetes Diet, Gestational Diabetes" 93: 821: 468: 327: 312: 245: 241: 1970:"Redox polymers in electrochemical systems: From methods of mediation to energy storage" 1158: 1072: 714:
Please help update this article to reflect recent events or newly available information.
225:
as well as a reduction in the short-term, potentially life-threatening complications of
1737: 1712: 1708: 1609:
The Pursuit of Noninvasive Glucose, 3rd Edition, by John L. Smith, Ph.D., available at
1381: 1354: 1330: 1303: 1262: 1235: 1231: 1207: 1180: 1053: 817: 292: 170: 1625:"Noninvasive glucose monitoring: comprehensive clinical results. Cygnus Research Team" 2035: 1995: 1665:. The National Institute of Diabetes and Digestive and Kidney Diseases. December 2008 1611:
http://www.mendosa.com/The%20Pursuit%20of%20Noninvasive%20Glucose%203rd%20Edition.pdf
1021: 874: 824:(sometimes known as GOx). Others use a similar reaction catalysed instead by another 663: 222: 1866:"Google Unveils Smart Contact Lens That Lets Diabetics Measure Their Glucose Levels" 1850: 1578: 467:: A consumable element, different for each meter, containing spots impregnated with 1017: 853: 796: 756: 565: 237: 198: 194: 182: 777: 299:, now part of Roche, who produce test strips called 'BM-test' for use in a meter. 1986: 1969: 1954: 1760: 1800: 893: 836: 600: 397: 315:
eventually persuaded acceptance. Prior to its discontinuation in July 2021, the
35: 1371: 772:
model UX5000, VX5200, and LX350 cell phones, as well as an add-on pack for the
1870: 1623:
Tamada JA, Garg S, Jovanovic L, Pitzer KR, Fermi S, Potts RO (November 1999).
1133: 865: 748: 732: 667: 528: 320: 1728: 1640: 1320: 1252: 1197: 857: 840: 572:. Meters have been combined with devices such as insulin injection devices, 551: 272: 197:. A small drop of blood, obtained from slightly piercing a fingertip with a 141: 1842: 1834: 1746: 1648: 1570: 1390: 1339: 1271: 1216: 323:
for the past 30 years, despite there being no such regulatory requirement.
540:
for measuring both unhealthy ketoacidosis and healthy nutritional ketosis.
17: 1302:
Han, Julia; Nichols, James H.; Rice, Mark; Klonoff, David C. (May 2020).
773: 671: 303: 626: 27:
Medical device for determining the concentration of glucose in the blood
2022: 781: 760: 537: 422: in this section. Unsourced material may be challenged and removed. 174: 1562: 240:
presented his first paper about the oxygen electrode, later named the
1155:"Diabetic Seniors – Informational Resource for Seniors with Diabetes" 1024:(high sugar), meters have been less useful. The primary problems are 825: 792: 765: 202: 1304:"The End of the Road for the YSI 2300 Analyzer: Where Do We Go Now?" 988:. Statements consisting only of original research should be removed. 1684: 564:: Some meters can transfer stored data, typically to a computer or 807: 625: 457:: The typical size is smaller than the palm of the hand. They are 383: 375: 284: 276: 250: 178: 784:. Similar systems have been tested for a longer time in Finland. 255:
CDC image showing the usage of a lancet and a blood glucose meter
1707:
Freckmann, G; Schmid, C; Baumstark, A; Rutschmann, M; Haug, C;
588: 953: 887: 690: 391: 208:
Since approximately 1980, a primary goal of the management of
29: 1404: 769: 977: 843:
derivative, which is oxidized to a blue polymer by the
816:
Many glucose meters employ the oxidation of glucose to
550:: Most meters include a memory to store test results, 1106: 1104: 266:
Home glucose monitoring was demonstrated to improve
151: 134: 60:. Unsourced material may be challenged and removed. 1236:"The YSI 2300 Analyzer Replacement Meeting Report" 1094: 1092: 776:Razr cell phone. In US, this limits providers to 173:for determining the approximate concentration of 1968:Yuan, Mengwai; Minteer, Shelley D. (June 2019). 835:The first-generation devices relied on the same 2042:Diabetes-related supplies and medical equipment 1355:"Glucose Measurement: Time for a Gold Standard" 317:YSI 2300 STAT PLUS Glucose and Lactate Analyzer 1894:"Google develops contact lens glucose monitor" 656:International Organization for Standardization 155:measure concentration of glucose in the blood 8: 1444: 1442: 1030:ratio of false positive and negative results 704:Parts of this article (those related to the 922:. Unsourced material may be challenged and 1717:Journal of Diabetes Science and Technology 1359:Journal of Diabetes Science and Technology 1308:Journal of Diabetes Science and Technology 1240:Journal of Diabetes Science and Technology 1185:Journal of Diabetes Science and Technology 140: 1985: 1736: 1473: 1471: 1380: 1370: 1329: 1319: 1261: 1251: 1206: 1196: 1004:Learn how and when to remove this message 942:Learn how and when to remove this message 438:Learn how and when to remove this message 120:Learn how and when to remove this message 1864:Lardinois, Frederic (January 16, 2014). 1523:. Diabetesincontrol.com. Archived from 1518:"Diabetes In Control : Newsletter" 1064: 368:computer systems for billing purposes. 291:to many health care professionals. In 131: 1481:. Abbott. 1 July 2005. Archived from 1450:"Unit of Measurement for Blood Sugar" 326:Home glucose testing was adopted for 7: 2057:Drugs developed by Hoffmann-La Roche 1892:Mendoza, Martha (January 16, 2014). 920:adding citations to reliable sources 420:adding citations to reliable sources 259:Another early glucose meter was the 58:adding citations to reliable sources 1974:Current Opinion in Electrochemistry 1943:Current Opinion in Electrochemistry 1916:"How Do Blood-Glucose Meters Work?" 1157:. Diabetes-wise.net. Archived from 1119:"Portable Meter To Aid Diabetics", 289:synonymous with the generic product 380:Glucose monitoring with glucometer 25: 1479:"Unit of Measure Reference Table" 1497:"Print Diabetes Self-Management" 958: 892: 695: 592:treatment. In the UK, where the 396: 189:(HBGM) performed by people with 34: 1663:"Continuous Glucose Monitoring" 1454:Top Blood Sugar Monitor Reviews 795:reported testing prototypes of 407:needs additional citations for 45:needs additional citations for 1781:"Dexcom Product Compatibility" 1134:"Insulin Pumpers UK: Glossary" 279:, and the Accu-Chek meter (by 1: 1823:Advanced Healthcare Materials 1353:Hagvik, Joakim (March 2007). 388:Glucose meter and test strips 1987:10.1016/j.coelec.2019.03.003 1955:10.1016/j.coelec.2017.06.007 1499:. Diabetesselfmanagement.com 576:, and cellular transmitters. 570:diabetes management software 1761:"Dexcom G4 product website" 984:the claims made and adding 705: 622:Continuous glucose monitors 613:Noninvasive glucose monitor 2078: 1372:10.1177/193229680700100205 1123:, November 5, 1981, p. A-6 1073:"Definition of GLUCOMETER" 1042:congenital hyperinsulinism 884:Meter use for hypoglycemia 873:can result from prolonged 637:Continuous glucose monitor 634: 610: 525:Glucose vs. plasma glucose 362:electronic medical records 2052:Medical testing equipment 2047:Physiological instruments 139: 1803:. Eworldwire. 2010-03-25 1729:10.1177/1932296815580160 1641:10.1001/jama.282.19.1839 1426:"Alternate site testing" 1321:10.1177/1932296819886603 1253:10.1177/1932296820911471 1198:10.1177/1932296816634356 1136:. Insulin-pumpers.org.uk 283:). Consequently, these 214:type 2 diabetes mellitus 187:blood glucose monitoring 1077:www.merriam-webster.com 1040:of hypoglycemia (e.g., 594:National Health Service 548:Memory and timestamping 536:in the blood to detect 356:Hospital glucose meters 219:long-term complications 165:also referred to as a " 1835:10.1002/adhm.201400504 813: 632: 495:Alternate site testing 489:Volume of blood sample 389: 381: 261:Ames Reflectance Meter 256: 1687:. Abbott Laboratories 1485:on February 19, 2011. 830:glucose dehydrogenase 811: 768:, an add-on pack for 629: 387: 379: 335:Non-meter test strips 254: 916:improve this section 708:section) need to be 681:Consensus Error Grid 534:beta-hydroxybutyrate 416:improve this article 54:improve this article 2062:American inventions 1697:An example of a CGM 660:ambient temperature 297:Boehringer Mannheim 2023:10.1039/C4RA16852F 1896:. Associated Press 1599:. 16 October 2006. 1456:. 11 February 2017 969:possibly contains 814: 644:interstitial fluid 633: 607:Noninvasive meters 390: 382: 302:In North America, 287:names have become 257: 1685:"FreeStyle Libre" 1563:10.1111/dme.12977 1551:Diabetic Medicine 1020:(low sugar) than 1014: 1013: 1006: 971:original research 952: 951: 944: 845:hydrogen peroxide 729: 728: 676:Clarke Error Grid 670:, dirt on meter, 448: 447: 440: 372:Test strip meters 191:diabetes mellitus 185:, including home 159: 158: 130: 129: 122: 104: 16:(Redirected from 2069: 2027: 2026: 2017:(18): 13570–78. 2006: 2000: 1999: 1989: 1965: 1959: 1958: 1938: 1932: 1931: 1929: 1927: 1918:. Archived from 1912: 1906: 1905: 1903: 1901: 1889: 1883: 1882: 1880: 1878: 1861: 1855: 1854: 1818: 1812: 1811: 1809: 1808: 1797: 1791: 1790: 1788: 1787: 1777: 1771: 1770: 1768: 1767: 1757: 1751: 1750: 1740: 1704: 1698: 1696: 1694: 1692: 1681: 1675: 1674: 1672: 1670: 1659: 1653: 1652: 1620: 1614: 1607: 1601: 1600: 1589: 1583: 1582: 1545: 1539: 1538: 1536: 1535: 1529: 1522: 1514: 1508: 1507: 1505: 1504: 1493: 1487: 1486: 1475: 1466: 1465: 1463: 1461: 1446: 1437: 1436: 1434: 1433: 1422: 1416: 1415: 1413: 1412: 1407:. Diabetes.co.uk 1401: 1395: 1394: 1384: 1374: 1350: 1344: 1343: 1333: 1323: 1299: 1293: 1292: 1290: 1282: 1276: 1275: 1265: 1255: 1227: 1221: 1220: 1210: 1200: 1176: 1170: 1169: 1167: 1166: 1151: 1145: 1144: 1142: 1141: 1130: 1124: 1121:Pittsburgh Press 1117: 1111: 1108: 1099: 1096: 1087: 1086: 1084: 1083: 1069: 1009: 1002: 998: 995: 989: 986:inline citations 962: 961: 954: 947: 940: 936: 933: 927: 896: 888: 744:Minimed Paradigm 724: 721: 715: 699: 698: 691: 514:United Kingdom. 501:Duration of test 443: 436: 432: 429: 423: 400: 392: 313:endocrinologists 268:glycemic control 144: 132: 125: 118: 114: 111: 105: 103: 62: 38: 30: 21: 2077: 2076: 2072: 2071: 2070: 2068: 2067: 2066: 2032: 2031: 2030: 2008: 2007: 2003: 1967: 1966: 1962: 1940: 1939: 1935: 1925: 1923: 1922:on 4 March 2016 1914: 1913: 1909: 1899: 1897: 1891: 1890: 1886: 1876: 1874: 1863: 1862: 1858: 1820: 1819: 1815: 1806: 1804: 1799: 1798: 1794: 1785: 1783: 1779: 1778: 1774: 1765: 1763: 1759: 1758: 1754: 1706: 1705: 1701: 1690: 1688: 1683: 1682: 1678: 1668: 1666: 1661: 1660: 1656: 1635:(19): 1839–44. 1622: 1621: 1617: 1608: 1604: 1591: 1590: 1586: 1547: 1546: 1542: 1533: 1531: 1527: 1520: 1516: 1515: 1511: 1502: 1500: 1495: 1494: 1490: 1477: 1476: 1469: 1459: 1457: 1448: 1447: 1440: 1431: 1429: 1428:. Accu-Chek.com 1424: 1423: 1419: 1410: 1408: 1403: 1402: 1398: 1352: 1351: 1347: 1301: 1300: 1296: 1288: 1284: 1283: 1279: 1232:Heinemann, Lutz 1229: 1228: 1224: 1178: 1177: 1173: 1164: 1162: 1153: 1152: 1148: 1139: 1137: 1132: 1131: 1127: 1118: 1114: 1109: 1102: 1097: 1090: 1081: 1079: 1071: 1070: 1066: 1062: 1050: 1010: 999: 993: 990: 975: 963: 959: 948: 937: 931: 928: 913: 897: 886: 822:glucose oxidase 806: 747:2016). It is a 725: 719: 716: 713: 700: 696: 689: 652: 639: 624: 615: 609: 585: 469:glucose oxidase 444: 433: 427: 424: 413: 401: 374: 358: 353: 351:Types of meters 337: 328:type 2 diabetes 246:glucose oxidase 242:Clark electrode 235: 210:type 1 diabetes 183:glucose testing 147: 126: 115: 109: 106: 69:"Glucose meter" 63: 61: 51: 39: 28: 23: 22: 15: 12: 11: 5: 2075: 2073: 2065: 2064: 2059: 2054: 2049: 2044: 2034: 2033: 2029: 2028: 2001: 1960: 1933: 1907: 1884: 1856: 1829:(6): 792–810. 1813: 1792: 1772: 1752: 1699: 1676: 1654: 1615: 1602: 1584: 1540: 1509: 1488: 1467: 1438: 1417: 1396: 1365:(2): 169–172. 1345: 1314:(3): 595–600. 1294: 1277: 1246:(3): 679–686. 1222: 1191:(4): 872–875. 1171: 1146: 1125: 1112: 1100: 1088: 1063: 1061: 1058: 1057: 1056: 1054:ISO/IEEE 11073 1049: 1046: 1012: 1011: 966: 964: 957: 950: 949: 900: 898: 891: 885: 882: 818:gluconolactone 805: 802: 797:contact lenses 791:In early 2014 727: 726: 703: 701: 694: 688: 685: 651: 648: 635:Main article: 623: 620: 611:Main article: 608: 605: 584: 581: 580: 579: 578: 577: 556: 555: 544: 543: 542: 541: 511: 510: 504: 498: 492: 485: 484: 483: 482: 473: 472: 462: 446: 445: 404: 402: 395: 373: 370: 357: 354: 352: 349: 336: 333: 234: 231: 171:medical device 163:glucose meter, 157: 156: 153: 149: 148: 145: 137: 136: 128: 127: 42: 40: 33: 26: 24: 14: 13: 10: 9: 6: 4: 3: 2: 2074: 2063: 2060: 2058: 2055: 2053: 2050: 2048: 2045: 2043: 2040: 2039: 2037: 2024: 2020: 2016: 2012: 2005: 2002: 1997: 1993: 1988: 1983: 1979: 1975: 1971: 1964: 1961: 1956: 1952: 1948: 1944: 1937: 1934: 1921: 1917: 1911: 1908: 1895: 1888: 1885: 1873: 1872: 1867: 1860: 1857: 1852: 1848: 1844: 1840: 1836: 1832: 1828: 1824: 1817: 1814: 1802: 1796: 1793: 1782: 1776: 1773: 1762: 1756: 1753: 1748: 1744: 1739: 1734: 1730: 1726: 1723:(4): 885–94. 1722: 1718: 1714: 1711:(July 2015). 1710: 1703: 1700: 1686: 1680: 1677: 1664: 1658: 1655: 1650: 1646: 1642: 1638: 1634: 1630: 1626: 1619: 1616: 1612: 1606: 1603: 1598: 1594: 1588: 1585: 1580: 1576: 1572: 1568: 1564: 1560: 1557:(5): 681–90. 1556: 1552: 1544: 1541: 1530:on 2015-09-23 1526: 1519: 1513: 1510: 1498: 1492: 1489: 1484: 1480: 1474: 1472: 1468: 1455: 1451: 1445: 1443: 1439: 1427: 1421: 1418: 1406: 1400: 1397: 1392: 1388: 1383: 1378: 1373: 1368: 1364: 1360: 1356: 1349: 1346: 1341: 1337: 1332: 1327: 1322: 1317: 1313: 1309: 1305: 1298: 1295: 1287: 1281: 1278: 1273: 1269: 1264: 1259: 1254: 1249: 1245: 1241: 1237: 1233: 1226: 1223: 1218: 1214: 1209: 1204: 1199: 1194: 1190: 1186: 1182: 1175: 1172: 1161:on 2014-11-08 1160: 1156: 1150: 1147: 1135: 1129: 1126: 1122: 1116: 1113: 1107: 1105: 1101: 1095: 1093: 1089: 1078: 1074: 1068: 1065: 1059: 1055: 1052: 1051: 1047: 1045: 1043: 1037: 1033: 1031: 1027: 1023: 1022:hyperglycemia 1019: 1008: 1005: 997: 987: 983: 979: 973: 972: 967:This article 965: 956: 955: 946: 943: 935: 925: 921: 917: 911: 910: 906: 901:This section 899: 895: 890: 889: 883: 881: 878: 876: 875:hyperglycemia 870: 867: 862: 859: 855: 849: 846: 842: 838: 833: 831: 827: 823: 820:catalyzed by 819: 810: 803: 801: 798: 794: 789: 785: 783: 779: 775: 771: 767: 762: 758: 753: 750: 745: 740: 736: 734: 723: 720:November 2017 711: 707: 702: 693: 692: 686: 684: 682: 677: 673: 669: 665: 664:ascorbic acid 661: 657: 649: 647: 645: 638: 628: 621: 619: 614: 606: 604: 602: 597: 595: 590: 582: 575: 571: 567: 563: 562:Data transfer 560: 559: 558: 557: 553: 549: 546: 545: 539: 535: 530: 526: 523: 522: 521: 520: 519: 515: 508: 505: 502: 499: 496: 493: 490: 487: 486: 480: 477: 476: 475: 474: 470: 466: 463: 460: 456: 453: 452: 451: 442: 439: 431: 428:November 2010 421: 417: 411: 410: 405:This section 403: 399: 394: 393: 386: 378: 371: 369: 367: 363: 355: 350: 348: 346: 341: 334: 332: 329: 324: 322: 318: 314: 309: 305: 300: 298: 294: 290: 286: 282: 278: 274: 269: 264: 262: 253: 249: 247: 243: 239: 232: 230: 228: 224: 223:hyperglycemia 220: 215: 211: 206: 204: 200: 196: 192: 188: 184: 180: 176: 172: 168: 164: 154: 150: 143: 138: 135:Glucose meter 133: 124: 121: 113: 110:November 2010 102: 99: 95: 92: 88: 85: 81: 78: 74: 71: â€“  70: 66: 65:Find sources: 59: 55: 49: 48: 43:This article 41: 37: 32: 31: 19: 2014: 2011:RSC Advances 2010: 2004: 1977: 1973: 1963: 1949:(1): 66–73. 1946: 1942: 1936: 1924:. Retrieved 1920:the original 1910: 1898:. Retrieved 1887: 1875:. Retrieved 1869: 1859: 1826: 1822: 1816: 1805:. Retrieved 1795: 1784:. Retrieved 1775: 1764:. Retrieved 1755: 1720: 1716: 1709:Heinemann, L 1702: 1689:. Retrieved 1679: 1667:. Retrieved 1657: 1632: 1628: 1618: 1605: 1596: 1587: 1554: 1550: 1543: 1532:. Retrieved 1525:the original 1512: 1501:. Retrieved 1491: 1483:the original 1458:. Retrieved 1453: 1430:. Retrieved 1420: 1409:. Retrieved 1399: 1362: 1358: 1348: 1311: 1307: 1297: 1280: 1243: 1239: 1230:Han, Julia; 1225: 1188: 1184: 1174: 1163:. Retrieved 1159:the original 1149: 1138:. Retrieved 1128: 1120: 1115: 1080:. Retrieved 1076: 1067: 1038: 1034: 1029: 1025: 1018:hypoglycemia 1015: 1000: 991: 968: 938: 929: 914:Please help 902: 879: 871: 863: 854:ferricyanide 850: 837:colorimetric 834: 815: 790: 786: 757:insulin pump 754: 741: 737: 730: 717: 709: 666:) in blood, 653: 640: 616: 598: 586: 566:mobile phone 561: 547: 524: 516: 512: 506: 500: 494: 488: 478: 464: 454: 449: 434: 425: 414:Please help 409:verification 406: 359: 342: 338: 325: 308:laboratories 301: 275:is owned by 265: 258: 238:Leland Clark 236: 227:hypoglycemia 207: 195:hypoglycemia 166: 162: 160: 116: 107: 97: 90: 83: 76: 64: 52:Please help 47:verification 44: 1926:28 November 1900:January 17, 1877:January 17, 1691:21 February 1669:21 February 866:coulometric 601:counterfeit 599:Batches of 552:timestamped 465:Test strips 321:glucometers 2036:Categories 1871:TechCrunch 1807:2014-03-13 1786:2016-01-30 1766:2016-01-30 1534:2014-03-13 1503:2014-03-13 1432:2018-07-20 1411:2014-03-13 1165:2014-03-13 1140:2014-03-13 1082:2020-10-10 1060:References 978:improve it 852:such as a 804:Technology 749:hypodermic 733:Cygnus Inc 668:hematocrit 529:hematocrit 366:laboratory 167:glucometer 80:newspapers 18:Glucometer 1996:108142625 1026:precision 982:verifying 903:does not 858:ferrocene 841:benzidine 461:-powered. 304:hospitals 273:trademark 169:", is a 1851:35508652 1843:25400274 1747:25872965 1649:10573275 1597:BioSpace 1579:19190997 1571:26443548 1391:19888402 1340:31711305 1272:32174135 1217:26902794 1048:See also 994:May 2023 932:May 2023 778:AT&T 774:Motorola 672:humidity 650:Accuracy 568:running 364:and the 1980:: 1–6. 1738:4525642 1382:2771464 1331:7576956 1263:7576944 1208:4928230 976:Please 924:removed 909:sources 856:ion, a 782:Verizon 761:wristop 710:updated 631:levels. 538:ketosis 507:Display 459:battery 293:Britain 233:History 177:in the 175:glucose 152:Purpose 94:scholar 1994:  1849:  1841:  1745:  1735:  1647:  1577:  1569:  1460:30 May 1389:  1379:  1338:  1328:  1291:. YSI. 1270:  1260:  1215:  1205:  826:enzyme 793:Google 766:iPhone 759:, and 706:Future 687:Future 479:Coding 203:mmol/L 199:lancet 96:  89:  82:  75:  67:  1992:S2CID 1847:S2CID 1575:S2CID 1528:(PDF) 1521:(PDF) 1289:(PDF) 285:brand 281:Roche 277:Bayer 221:from 179:blood 101:JSTOR 87:books 1928:2012 1902:2014 1879:2014 1839:PMID 1743:PMID 1693:2016 1671:2016 1645:PMID 1629:JAMA 1567:PMID 1462:2023 1387:PMID 1336:PMID 1268:PMID 1213:PMID 1028:and 907:any 905:cite 780:and 589:US$ 583:Cost 574:PDAs 455:Size 212:and 193:or 73:news 2019:doi 1982:doi 1951:doi 1831:doi 1733:PMC 1725:doi 1637:doi 1633:282 1559:doi 1377:PMC 1367:doi 1326:PMC 1316:doi 1258:PMC 1248:doi 1203:PMC 1193:doi 980:by 918:by 418:by 345:NHS 56:by 2038:: 2013:. 1990:. 1978:15 1976:. 1972:. 1945:. 1868:. 1845:. 1837:. 1825:. 1741:. 1731:. 1719:. 1715:. 1643:. 1631:. 1627:. 1595:. 1573:. 1565:. 1555:33 1553:. 1470:^ 1452:. 1441:^ 1385:. 1375:. 1361:. 1357:. 1334:. 1324:. 1312:14 1310:. 1306:. 1266:. 1256:. 1244:14 1242:. 1238:. 1211:. 1201:. 1189:10 1187:. 1183:. 1103:^ 1091:^ 1075:. 877:. 828:, 770:LG 229:. 205:. 161:A 2025:. 2021:: 2015:5 1998:. 1984:: 1957:. 1953:: 1947:5 1930:. 1904:. 1881:. 1853:. 1833:: 1827:4 1810:. 1789:. 1769:. 1749:. 1727:: 1721:9 1695:. 1673:. 1651:. 1639:: 1613:. 1581:. 1561:: 1537:. 1506:. 1464:. 1435:. 1414:. 1393:. 1369:: 1363:1 1342:. 1318:: 1274:. 1250:: 1219:. 1195:: 1168:. 1143:. 1085:. 1007:) 1001:( 996:) 992:( 974:. 945:) 939:( 934:) 930:( 926:. 912:. 722:) 718:( 712:. 441:) 435:( 430:) 426:( 412:. 123:) 117:( 112:) 108:( 98:· 91:· 84:· 77:· 50:. 20:)

Index

Glucometer

verification
improve this article
adding citations to reliable sources
"Glucose meter"
news
newspapers
books
scholar
JSTOR
Learn how and when to remove this message

medical device
glucose
blood
glucose testing
blood glucose monitoring
diabetes mellitus
hypoglycemia
lancet
mmol/L
type 1 diabetes
type 2 diabetes mellitus
long-term complications
hyperglycemia
hypoglycemia
Leland Clark
Clark electrode
glucose oxidase

Text is available under the Creative Commons Attribution-ShareAlike License. Additional terms may apply.

↑