Clover (detector)

Source 📝

83:. Large volume high purity single crystals of Ge can be expensive. By mounting four smaller crystals in a common cryostat a detector of a given volume can be created at a reduced cost. In addition, the individual smaller Ge crystals present a smaller solid angle than a large volume Ge detector thus significantly reducing the effects of 63:, a gamma ray may interact with two (or possibly more) crystals resulting in the energy (and thus the liberated charge) being shared by the crystals. In this case, a process known as add-back, where the charge collected by each of the crystals is summed, can be used to determine the energy of the incident gamma ray. 50:
The clover is the first composite Ge detector. It remains widely used in the detectors of particle accelerators, where multiple clover modules form an array all around the target to capture the rays. More complex composite detectors, such as the 7-element hexagonal cluster detector used on the
90:
A clover detector can also be used to determine the electric or magnetic nature of the incident photons (e.g. if the gamma ray is an electric quadrupole or a magnetic dipole) as the Compton scattering process for these two types of radiation is different. This is called
70:
using a fence of BGO detectors. A gamma ray can escape the Clover array via Compton scattering, causing incorrectly low charge values. A BGO detector can detect the escaped ray and have the computer ignore the wrong reading from the Clover.
59:
A gamma ray may interact with a single Ge crystal and deposit its full energy. The resulting charge collected will then be proportional to this energy. However, through the process of
87:
on the resulting spectra. Doppler broadening is further reduced using techniques that exploit the individual readings from the smaller crystals.
168:
Eberth, J.; Simpson, J. (2008). "From Ge(Li) detectors to gamma-ray tracking arrays–50 years of gamma spectroscopy with germanium detectors".
121:
DuchĂȘne, G.; Beck, F.A.; Twin, P.J.; de France, G.; Curien, D.; Han, L.; Beausang, C.W.; Bentley, M.A.; Nolan, P.J.; Simpson, J. (1999).
219:
Jones, P.M. (August 1985). "Calibration of the new composite "clover" detector as a Compton polarimeter for the EUROGAM array".
257: 127:
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
277: 79:
There are a number of advantages offered by using clover detectors as opposed to the more conventional single crystal
17: 272: 92: 228: 177: 134: 32: 67: 201: 84: 80: 60: 28: 193: 150: 236: 185: 142: 44: 232: 181: 138: 146: 266: 240: 205: 189: 197: 154: 36: 40: 122: 39:(Ge) crystals each machined to shape and mounted in a common 123:"The Clover: a new generation of composite Ge detectors" 116: 114: 112: 110: 108: 66:In addition to add-back, the Clover also uses 8: 170:Progress in Particle and Nuclear Physics 104: 258:CLover Array for Radioactive ION beams 7: 221:Nucl. Instrum. Methods Phys. Res. A 51:Euroball, offer even better data. 14: 43:to form a structure resembling a 1: 147:10.1016/S0168-9002(99)00277-6 241:10.1016/0168-9002(95)00246-4 31:that consists of 4 coaxial 294: 190:10.1016/j.ppnp.2007.09.001 15: 18:Clover (disambiguation) 16:For other uses, see 233:1995NIMPA.362..556J 182:2008PrPNP..60..283E 139:1999NIMPA.432...90D 93:Compton polarimetry 81:germanium detectors 68:Compton suppression 278:Particle detectors 85:Doppler broadening 61:Compton scattering 29:gamma-ray detector 285: 245: 244: 227:(2–3): 556–560. 216: 210: 209: 165: 159: 158: 118: 45:four-leaf clover 293: 292: 288: 287: 286: 284: 283: 282: 263: 262: 254: 249: 248: 218: 217: 213: 167: 166: 162: 120: 119: 106: 101: 77: 57: 25:clover detector 21: 12: 11: 5: 291: 289: 281: 280: 275: 265: 264: 261: 260: 253: 252:External links 250: 247: 246: 211: 176:(2): 283–337. 160: 103: 102: 100: 97: 76: 73: 56: 53: 13: 10: 9: 6: 4: 3: 2: 290: 279: 276: 274: 273:Spectrometers 271: 270: 268: 259: 256: 255: 251: 242: 238: 234: 230: 226: 222: 215: 212: 207: 203: 199: 195: 191: 187: 183: 179: 175: 171: 164: 161: 156: 152: 148: 144: 140: 136: 133:(1): 90–110. 132: 128: 124: 117: 115: 113: 111: 109: 105: 98: 96: 94: 88: 86: 82: 74: 72: 69: 64: 62: 54: 52: 48: 46: 42: 38: 34: 30: 26: 19: 224: 220: 214: 173: 169: 163: 130: 126: 89: 78: 65: 58: 49: 35:high purity 24: 22: 267:Categories 99:References 75:Advantages 206:121502646 198:0146-6410 155:0168-9002 55:Operation 37:germanium 41:cryostat 229:Bibcode 178:Bibcode 135:Bibcode 204:  196:  153:  33:N-type 202:S2CID 27:is a 194:ISSN 151:ISSN 237:doi 225:362 186:doi 143:doi 131:432 269:: 235:. 223:. 200:. 192:. 184:. 174:60 172:. 149:. 141:. 129:. 125:. 107:^ 95:. 47:. 23:A 243:. 239:: 231:: 208:. 188:: 180:: 157:. 145:: 137:: 20:.

Index

Clover (disambiguation)
gamma-ray detector
N-type
germanium
cryostat
four-leaf clover
Compton scattering
Compton suppression
germanium detectors
Doppler broadening
Compton polarimetry





"The Clover: a new generation of composite Ge detectors"
Bibcode
1999NIMPA.432...90D
doi
10.1016/S0168-9002(99)00277-6
ISSN
0168-9002
Bibcode
2008PrPNP..60..283E
doi
10.1016/j.ppnp.2007.09.001
ISSN
0146-6410
S2CID

Text is available under the Creative Commons Attribution-ShareAlike License. Additional terms may apply.

↑