Knowledge (XXG)

Fourier-transform infrared spectroscopy

Source 📝

646:, corresponding to a maximum OPD of 10 m. The point in the interferogram corresponding to zero path difference has to be identified, commonly by assuming it is where the maximum signal occurs. This so-called centerburst is not always symmetrical in real world spectrometers so a phase correction may have to be calculated. The interferogram signal decays as the path difference increases, the rate of decay being inversely related to the width of features in the spectrum. If the OPD is not large enough to allow the interferogram signal to decay to a negligible level there will be unwanted oscillations or sidelobes associated with the features in the resulting spectrum. To reduce these sidelobes the interferogram is usually multiplied by a function that approaches zero at the maximum OPD. This so-called 836:). These detectors operate at ambient temperatures and provide adequate sensitivity for most routine applications. To achieve the best sensitivity the time for a scan is typically a few seconds. Cooled photoelectric detectors are employed for situations requiring higher sensitivity or faster response. Liquid nitrogen cooled mercury cadmium telluride (MCT) detectors are the most widely used in the mid-IR. With these detectors an interferogram can be measured in as little as 10 milliseconds. Uncooled indium gallium arsenide photodiodes or DTGS are the usual choices in near-IR systems. Very sensitive liquid-helium-cooled silicon or germanium bolometers are used in the far-IR where both sources and beamsplitters are inefficient. 501:, any energy at shorter wavelengths would be interpreted as coming from longer wavelengths and so has to be minimized optically or electronically. The spectral resolution, i.e. the separation between wavelengths that can be distinguished, is determined by the maximum OPD. The wavelengths used in calculating the Fourier transform are such that an exact number of wavelengths fit into the length of the interferogram from zero to the maximum OPD as this makes their contributions orthogonal. This results in a spectrum with points separated by equal frequency intervals. 423:
to vary the path of one beam. In this arrangement the moving mirror must not tilt or wobble as this would affect how the beams overlap as they recombine. Some systems incorporate a compensating mechanism that automatically adjusts the orientation of one mirror to maintain the alignment. Arrangements that avoid this problem include using cube corner reflectors instead of plane mirrors as these have the property of returning any incident beam in a parallel direction regardless of orientation.
469: 402:. Ideally 50% of the light is refracted towards the fixed mirror and 50% is transmitted towards the moving mirror. Light is reflected from the two mirrors back to the beam splitter and some fraction of the original light passes into the sample compartment. There, the light is focused on the sample. On leaving the sample compartment the light is refocused on to the detector. The difference in optical path length between the two arms to the interferometer is known as the 819:). The output is similar to a blackbody. Shorter wavelengths of the near-IR, 1−2.5 μm (10,000–4,000 cm), require a higher temperature source, typically a tungsten-halogen lamp. The long wavelength output of these is limited to about 5 μm (2,000 cm) by the absorption of the quartz envelope. For the far-IR, especially at wavelengths beyond 50 μm (200 cm) a mercury discharge lamp gives higher output than a thermal source. 88: 427: 934:
precision optical and mechanical components had to be solved. A wide range of instruments are now available commercially. Although instrument design has become more sophisticated, the basic principles remain the same. Nowadays, the moving mirror of the interferometer moves at a constant velocity, and sampling of the interferogram is triggered by finding zero-crossings in the fringes of a secondary interferometer lit by a
2410: 163: 379: 320: 110: 1022:. The images contain a spectrum for each pixel and can be viewed as maps showing the intensity at any wavelength or combination of wavelengths. This allows the distribution of different chemical species within the sample to be seen. This technique has been applied in various biological applications including the analysis of tissue sections as an alternative to conventional 845: 2422: 1247: 702:
the collimated beam in the interferometer. This is because convergent rays are modulated at different frequencies as the path difference is varied. Such an aperture is called a Jacquinot stop. For a given resolution and wavelength this circular aperture allows more light through than a slit, resulting in a higher signal-to-noise ratio.
1001:
and proteins in hydrophobic membrane environments. Studies show the ability of FTIR to directly determine the polarity at a given site along the backbone of a transmembrane protein. The bond features involved with various organic and inorganic nanomaterials and their quantitative analysis can be done
710:
Another minor advantage is less sensitivity to stray light, that is radiation of one wavelength appearing at another wavelength in the spectrum. In dispersive instruments, this is the result of imperfections in the diffraction gratings and accidental reflections. In FT instruments there is no direct
705:
The wavelength accuracy or Connes' advantage. The wavelength scale is calibrated by a laser beam of known wavelength that passes through the interferometer. This is much more stable and accurate than in dispersive instruments where the scale depends on the mechanical movement of diffraction gratings.
701:
has entrance and exit slits which restrict the amount of light that passes through it. The interferometer throughput is determined only by the diameter of the collimated beam coming from the source. Although no slits are needed, FTIR spectrometers do require an aperture to restrict the convergence of
442:
into one of the beams. Increasing the thickness of KBr in the beam increases the optical path because the refractive index is higher than that of air. One limitation of this approach is that the variation of refractive index over the wavelength range limits the accuracy of the wavelength calibration.
731:
in cm is equal to the reciprocal of the maximal retardation in cm. Thus a 4 cm resolution will be obtained if the maximal retardation is 0.25 cm; this is typical of the cheaper FTIR instruments. Much higher resolution can be obtained by increasing the maximal retardation. This is not easy,
464:
to measure the IR signal each time the laser signal passes through zero. Alternatively, the laser and IR signals can be measured synchronously at smaller intervals with the IR signal at points corresponding to the laser signal zero crossing being determined by interpolation. This approach allows the
1683:
Deepty, M., Ch Srinivas, E. Ranjith Kumar, N. Krisha Mohan, C. L. Prajapat, TV Chandrasekhar Rao, Sher Singh Meena, Amit Kumar Verma, and D. L. Sastry. "XRD, EDX, FTIR and ESR spectroscopic studies of co-precipitated Mn–substituted Zn–ferrite nanoparticles." Ceramics International 45, no. 6 (2019):
1054:
The speed of FTIR allows spectra to be obtained from compounds as they are separated by a gas chromatograph. However this technique is little used compared to GC-MS (gas chromatography-mass spectrometry) which is more sensitive. The GC-IR method is particularly useful for identifying isomers, which
912:
The first FTIR spectrometers were developed for far-infrared range. The reason for this has to do with the mechanical tolerance needed for good optical performance, which is related to the wavelength of the light being used. For the relatively long wavelengths of the far infrared, ~10 μm tolerances
852:
An ideal beam-splitter transmits and reflects 50% of the incident radiation. However, as any material has a limited range of optical transmittance, several beam-splitters may be used interchangeably to cover a wide spectral range. For the mid-IR region the beamsplitter is usually made of KBr with a
422:
Commercial spectrometers use Michelson interferometers with a variety of scanning mechanisms to generate the path difference. Common to all these arrangements is the need to ensure that the two beams recombine exactly as the system scans. The simplest systems have a plane mirror that moves linearly
865:
is the usual material for the near-IR, being both harder and less sensitive to moisture than KBr but cannot be used beyond about 8 μm (1,200 cm). In a simple Michelson interferometer one beam passes twice through the beamsplitter but the other passes through only once. To correct for this an
434:
Systems where the path difference is generated by a rotary movement have proved very successful. One common system incorporates a pair of parallel mirrors in one beam that can be rotated to vary the path without displacing the returning beam. Another is the double pendulum design where the path in
455:
The interferogram is converted to a spectrum by Fourier transformation. This requires it to be stored in digital form as a series of values at equal intervals of the path difference between the two beams. To measure the path difference a laser beam is sent through the interferometer, generating a
451:
The interferogram has to be measured from zero path difference to a maximum length that depends on the resolution required. In practice the scan can be on either side of zero resulting in a double-sided interferogram. Mechanical design limitations may mean that for the highest resolution the scan
197:
beam of light (a beam composed of only a single wavelength) at the sample, this technique shines a beam containing many frequencies of light at once and measures how much of that beam is absorbed by the sample. Next, the beam is modified to contain a different combination of frequencies, giving a
769:
atom into its two components by using his interferometer. A spectrometer with 0.001 cm resolution is now available commercially. The throughput advantage is important for high-resolution FTIR, as the monochromator in a dispersive instrument with the same resolution would have very narrow
1075:
FTIR analysis is used to determine water content in fairly thin plastic and composite parts, more commonly in the laboratory setting. Such FTIR methods have long been used for plastics, and became extended for composite materials in 2018, when the method was introduced by Krauklis, Gagani and
933:
it became possible to have a computer dedicated to controlling the spectrometer, collecting the data, doing the Fourier transform and presenting the spectrum. This provided the impetus for the development of FTIR spectrometers for the rock-salt region. The problems of manufacturing ultra-high
984:
FTIR can be used in all applications where a dispersive spectrometer was used in the past (see external links). In addition, the improved sensitivity and speed have opened up new areas of application. Spectra can be measured in situations where very little energy reaches the detector aaurier
799: 476:
The result of Fourier transformation is a spectrum of the signal at a series of discrete wavelengths. The range of wavelengths that can be used in the calculation is limited by the separation of the data points in the interferogram. The shortest wavelength that can be recognized is twice the
418:
at all wavelengths, followed by series of "wiggles". The position of zero retardation is determined accurately by finding the point of maximum intensity in the interferogram. When a sample is present the background interferogram is modulated by the presence of absorption bands in the sample.
657:
the number of points in the interferogram has to equal a power of two. A string of zeroes may be added to the measured interferogram to achieve this. More zeroes may be added in a process called zero filling to improve the appearance of the final spectrum although there is no improvement in
638:. This is the spectral resolution in the sense that the value at one point is independent of the values at adjacent points. Most instruments can be operated at different resolutions by choosing different OPD's. Instruments for routine analyses typically have a best resolution of around 1525:
Baker, Matthew J.; Trevisan, Júlio; Bassan, Paul; Bhargava, Rohit; Butler, Holly J.; Dorling, Konrad M.; Fielden, Peter R.; Fogarty, Simon W.; Fullwood, Nigel J.; Heys, Kelly A.; Hughes, Caryn; Lasch, Peter; Martin-Hirsch, Pierre L.; Obinaju, Blessing; Sockalingum, Ganesh D. (2014).
293:
were well-known, but considerable technical difficulties had to be overcome before a commercial instrument could be built. Also an electronic computer was needed to perform the required Fourier transform, and this only became practicable with the advent of
938:. In modern FTIR systems the constant mirror velocity is not strictly required, as long as the laser fringes and the original interferogram are recorded simultaneously with higher sampling rate and then re-interpolated on a constant grid, as pioneered by 302:, which became available in 1965. Digilab pioneered the world's first commercial FTIR spectrometer (Model FTS-14) in 1969. Digilab FTIRs are now a part of Agilent technologies's molecular product line after Agilent acquired spectroscopy business from 212:
As mentioned, computer processing is required to turn the raw data (light absorption for each mirror position) into the desired result (light absorption for each wavelength). The processing required turns out to be a common algorithm called the
883:(ATR) is one accessory of FTIR spectrophotometer to measure surface properties of solid or thin film samples rather than their bulk properties. Generally, ATR has a penetration depth of around 1 or 2 micrometers depending on sample conditions. 736:
mirrors in place of the flat mirrors is helpful, as an outgoing ray from a corner-cube mirror is parallel to the incoming ray, regardless of the orientation of the mirror about axes perpendicular to the axis of the light beam. In 1966
1055:
by their nature have identical masses. Liquid chromatography fractions are more difficult because of the solvent present. One notable exception is to measure chain branching as a function of molecular size in polyethylene using
414:. The form of the interferogram when no sample is present depends on factors such as the variation of source intensity and splitter efficiency with wavelength. This results in a maximum at zero retardation, when there is 205:—a certain configuration of mirrors, one of which is moved by a motor. As this mirror moves, each wavelength of light in the beam is periodically blocked, transmitted, blocked, transmitted, by the interferometer, due to 831:
that respond to changes in temperature as the intensity of IR radiation falling on them varies. The sensitive elements in these detectors are either deuterated triglycine sulfate (DTGS) or lithium tantalate
986: 198:
second data point. This process is rapidly repeated many times over a short time span. Afterwards, a computer takes all this data and works backward to infer what the absorption is at each wavelength.
694:. Alternatively, it allows a shorter scan-time for a given resolution. In practice multiple scans are often averaged, increasing the signal-to-noise ratio by the square root of the number of scans. 1674:"Structural, functional and magnetic ordering modifications in graphene oxide and graphite by 100 MeV gold ion irradiation". Vacuum. 182: 109700. 2020-12-01. doi:10.1016/j.vacuum.2020.109700 1067:
Measuring the gas evolved as a material is heated allows qualitative identification of the species to complement the purely quantitative information provided by measuring the weight loss.
337: 127: 2241: 891:
The interferogram in practice consists of a set of intensities measured for discrete values of retardation. The difference between successive retardation values is constant. Thus, a
217:. The Fourier transform converts one domain (in this case displacement of the mirror in cm) into its inverse domain (wavenumbers in cm). The raw data is called an "interferogram". 861:
are sometimes used to extend the range to about 50 μm (200 cm). ZnSe is an alternative where moisture vapor can be a problem but is limited to about 20μm (500 cm). CaF
59:
of a solid, liquid, or gas. An FTIR spectrometer simultaneously collects high-resolution spectral data over a wide spectral range. This confers a significant advantage over a
1046:
and allows for performing broadband spectroscopy on materials in ultra-small quantities (single viruses and protein complexes) and with 10 to 20 nm spatial resolution.
1914: 1738:
Amenabar, Iban; Poly, Simon; Nuansing, Wiwat; Hubrich, Elmar H.; Govyadinov, Alexander A.; Huth, Florian; Krutokhvostov, Roman; Zhang, Lianbing; Knez, Mato (2013-12-04).
182:, etc.) is to measure how much light a sample absorbs at each wavelength. The most straightforward way to do this, the "dispersive spectroscopy" technique, is to shine a 811:
FTIR spectrometers are mostly used for measurements in the mid and near IR regions. For the mid-IR region, 2−25 μm (5,000–400 cm), the most common source is a
209:. Different wavelengths are modulated at different rates, so that at each moment or mirror position the beam coming out of the interferometer has a different spectrum. 201:
The beam described above is generated by starting with a broadband light source—one containing the full spectrum of wavelengths to be measured. The light shines into a
1014:
allows samples to be observed and spectra measured from regions as small as 5 microns across. Images can be generated by combining a microscope with linear or 2-D
166:
An FTIR interferogram. The central peak is at the ZPD position ("zero path difference" or zero retardation), where the maximal amount of light passes through the
914: 2132: 866:
additional compensator plate of equal thickness is incorporated. Far-IR beamsplitters are mostly based on polymer films and cover a limited wavelength range.
2065: 2010: 1979: 1974: 913:
are adequate, whereas for the rock-salt region tolerances have to be better than 1 μm. A typical instrument was the cube interferometer developed at the
257:), which becomes opaque at wavelengths longer than about 15 μm; this spectral region became known as the rock-salt region. Later instruments used 2347: 2165: 2027: 1217: 265:
50 μm (200 cm). The region beyond 50 μm (200 cm) became known as the far-infrared region; at very long wavelengths it merges into the
2296: 2115: 1959: 241:
range 4,000 cm to 660 cm). The lower wavelength limit was chosen to encompass the highest known vibration frequency due to a fundamental
2236: 2038: 1480:"Investigating the impact of spectral data pre-processing to assess honey botanical origin through Fourier transform infrared spectroscopy (FTIR)" 1076:
Echtermeyer. FTIR method uses the maxima of the absorbance band at about 5,200 cm−1 which correlates with the true water content in the material.
678:
for a given scan-time for observations limited by a fixed detector noise contribution (typically in the thermal infrared spectral region where a
2182: 2160: 1907: 1039: 52: 2248: 2170: 1182: 1148: 2000: 1632:
Brielle, Esther S.; Arkin, Isaiah T. (2018). "Site-Specific Hydrogen Exchange in a Membrane Environment Analyzed by Infrared Spectroscopy".
187: 2105: 2050: 56: 31: 1877: 465:
use of analog-to-digital converters that are more accurate and precise than converters that can be triggered, resulting in lower noise.
786:
people measure infrared absorption and emission spectra, i.e. why and how substances absorb and emit infrared light, see the article:
2332: 2084: 1900: 410:(OPD). An interferogram is obtained by varying the retardation and recording the signal from the detector for various values of the 359: 149: 723:(FT) inverts the dimension, so the FT of the interferogram belongs in the reciprocal length dimension(), that is the dimension of 456:
sinusoidal signal where the separation between successive maxima is equal to the wavelength of the laser (typically a 633 nm
2337: 2155: 179: 2448: 2352: 2322: 2253: 2187: 1103: 273:
to replace the prisms as dispersing elements, since salt crystals are opaque in this region. More sensitive detectors than the
186:
light beam at a sample, measure how much of the light is absorbed, and repeat for each different wavelength. (This is how some
2458: 2281: 2072: 1969: 1097: 683: 341: 131: 1388:
Chamberain, J.; Gibbs, J.E.; Gebbie, H.E. (1969). "The determination of refractive index spectra by fourier spectrometry".
2079: 1984: 1056: 853:
germanium-based coating that makes it semi-reflective. KBr absorbs strongly at wavelengths beyond 25 μm (400 cm) so
706:
In practice, the accuracy is limited by the divergence of the beam in the interferometer which depends on the resolution.
2213: 2060: 1949: 1248:"Agilent Technologies completes acquisition of Varian, Inc., marking historic milestone for two Silicon Valley pioneers" 880: 875: 92: 2369: 1478:
Tsagkaris, A. S.; Bechynska, K.; Ntakoulas, D. D.; Pasias, I. N.; Weller, P.; Proestos, C.; Hajslova, J. (2023-06-01).
2208: 2177: 2110: 1315:
Connes, J.; Connes, P. (1966). "Near-Infrared Planetary Spectra by Fourier Spectroscopy. I. Instruments and Results".
674:. This arises from the fact that information from all wavelengths is collected simultaneously. It results in a higher 461: 921:. It used a stepper motor to drive the moving mirror, recording the detector response after each step was completed. 634:
The separation is the inverse of the maximum OPD. For example, a maximum OPD of 2 cm results in a separation of
1697:
Scoble, Laura; Ussher, Simon J.; Fitzsimons, Mark F.; Ansell, Lauren; Craven, Matthew; Fyfe, Ralph M. (2024-02-01).
472:
Values of the interferogram at times corresponding to zero crossings of the laser signal are found by interpolation.
2359: 2301: 2150: 2022: 1085: 955: 892: 1699:"Optimisation of classification methods to differentiate morphologically-similar pollen grains from FT-IR spectra" 330: 120: 2453: 2385: 2364: 2127: 2005: 1883: 1865: 1805:"Near-Infrared Spectroscopic Method for Monitoring Water Content in Epoxy Resins and Fiber-Reinforced Composites" 415: 650:
reduces the amplitude of any sidelobes and also the noise level at the expense of some reduction in resolution.
968:
of fundamental vibrations can be observed in this region. It is used mainly in industrial applications such as
754: 387: 373: 290: 202: 167: 48: 666:
There are three principal advantages for an FT spectrometer compared to a scanning (dispersive) spectrometer.
477:
separation between these data points. For example, with one point per wavelength of a HeNe reference laser at
2426: 193:
Fourier-transform spectroscopy is a less intuitive way to obtain the same information. Rather than shining a
2258: 1954: 671: 175: 1342:
Smith, D.R.; Morgan, R.L.; Loewenstein, E.V. (1968). "Comparison of the Radiance of Far-Infrared Sources".
1088: – Type of Fourier transform in discrete mathematics − for computing periodicity in evenly spaced data 289:
in this region. Far-infrared spectrophotometers were cumbersome, slow and expensive. The advantages of the
2045: 1225: 896: 746: 654: 1038:
The spatial resolution of FTIR can be further improved below the micrometer scale by integrating it into
935: 697:
The throughput or Jacquinot's advantage. This results from the fact that in a dispersive instrument, the
457: 237:
Infracord produced in 1957. This instrument covered the wavelength range from 2.5 μm to 15 μm (
2414: 2286: 2017: 1931: 1887: 828: 787: 675: 230: 76: 1026:, examining the homogeneity of pharmaceutical tablets, and for differentiating morphologically-similar 711:
equivalent as the apparent wavelength is determined by the modulation frequency in the interferometer.
17: 1698: 1479: 960:
The near-infrared region spans the wavelength range between the rock-salt region and the start of the
942:. This confers very high wavenumber accuracy on the resulting infrared spectrum and avoids wavenumber 468: 1816: 1751: 1397: 1281: 2342: 2055: 1964: 1374: 728: 286: 270: 246: 242: 60: 798: 2390: 2327: 2306: 2122: 2100: 2033: 1944: 1657: 1507: 1460: 1272:
Brault, James W. (1996). "New Approach to high-precision Fourier-transform spectrometer design".
1425:"Renal geology (quantitative renal stone analysis) by 'Fourier transform infrared spectroscopy'" 1059:, which is possible using chlorinated solvents that have no absorption in the area in question. 2291: 2218: 2192: 1844: 1785: 1767: 1740:"Structural analysis and mapping of individual protein complexes by infrared nanospectroscopy" 1720: 1649: 1614: 1565: 1547: 1499: 1452: 1444: 1297: 1178: 1144: 1091: 1015: 720: 439: 258: 226: 214: 206: 81: 1872: 658:
resolution. Alternatively, interpolation after the Fourier transform gives a similar result.
1834: 1824: 1775: 1759: 1710: 1641: 1604: 1596: 1555: 1539: 1491: 1436: 1405: 1351: 1324: 1289: 973: 961: 858: 782:
FTIR is a method of measuring infrared absorption and emission spectra. For a discussion of
733: 250: 1094: – Mathematical transform that expresses a function of time as a function of frequency 426: 87: 1140: 1106: – Periodicity computation method − for computing periodicity in unevenly spaced data 969: 939: 812: 254: 438:
A quite different approach involves moving a wedge of an IR-transparent material such as
1820: 1755: 1401: 1285: 1839: 1804: 1780: 1739: 1609: 1584: 1560: 1527: 1023: 854: 278: 262: 84:(a mathematical process) is required to convert the raw data into the actual spectrum. 2442: 1511: 1409: 998: 930: 918: 771: 738: 698: 679: 399: 194: 183: 1661: 1018:
detectors. The spatial resolution can approach 5 microns with tens of thousands of
430:
Interferometer schematics where the path difference is generated by a rotary motion.
269:
region. Measurements in the far infrared needed the development of accurately ruled
1923: 1715: 1464: 378: 303: 295: 282: 234: 63: 411: 162: 1685: 1134: 1645: 943: 647: 390:
adapted for FTIR, light from the polychromatic infrared source, approximately a
319: 109: 1495: 530:
cycles, respectively, in the interferogram. The corresponding frequencies are ν
1440: 1011: 724: 395: 391: 238: 67: 1771: 1724: 1551: 1503: 1448: 732:
as the moving mirror must travel in a near-perfect straight line. The use of
1585:"Environment Polarity in Proteins Mapped Noninvasively by FTIR Spectroscopy" 1424: 1043: 965: 435:
one arm of the interferometer increases as the path in the other decreases.
274: 266: 1848: 1789: 1653: 1618: 1569: 1543: 1456: 1355: 1328: 1301: 1177:(Paperback ed.). New York: Oxford University Press. pp. 173–178. 844: 277:
were required because of the low energy of the radiation. One such was the
1528:"Using Fourier transform IR spectroscopy to analyze biological materials" 1293: 766: 498: 45: 1829: 1763: 758: 344: in this section. Unsourced material may be challenged and removed. 134: in this section. Unsourced material may be challenged and removed. 1600: 1027: 1019: 816: 382:
Schematic diagram of a Michelson interferometer, configured for FTIR
815:
element heated to about 1,200 K (930 °C; 1,700 °F) (
843: 797: 742: 690:
resolution elements, this increase is equal to the square root of
642:, while spectrometers have been built with resolutions as high as 467: 425: 377: 299: 161: 86: 1218:"Agilent Technologies to acquire Varian, Inc. for $ 1.5 Billion" 848:
Simple interferometer with a beam-splitter and compensator plate
1896: 1892: 313: 103: 1583:
Manor, Joshua; Feldblum, Esther S.; Arkin, Isaiah T. (2012).
989:, chemistry, materials, botany and biology research fields. 1803:
Krauklis, A. E.; Gagani, A. I.; Echtermeyer, A. T. (2018).
802:
FTIR setup. The sample is placed right before the detector.
281:. An additional issue is the need to exclude atmospheric 1100: – Spectroscopy based on time- or space-domain data 1063:
TG-IR (thermogravimetric analysis-infrared spectrometry)
1175:
Principles of materials characterization and metrology
1071:
Water content determination in plastics and composites
1034:
Nanoscale and spectroscopy below the diffraction limit
261:
prisms to extend the range to 25 μm (400 cm) and
27:
Technique to analyze the infrared spectrum of matter
2378: 2315: 2274: 2267: 2229: 2201: 2143: 2093: 1993: 1930: 719:The interferogram belongs in the length dimension. 245:. The upper limit was imposed by the fact that the 1880:Properties of many salt crystals and useful links. 452:runs to the maximum OPD on one side of zero only. 66:, which measures intensity over a narrow range of 30:"FTIR" redirects here. The term may also refer to 1873:Spectroscopy, part 2 by Dudley Williams, page 81 1133:Griffiths, P.; de Hasseth, J. A. (18 May 2007). 1042:platform. The corresponding technique is called 1198:"The Infracord double-beam spectrophotometer". 1686:https://doi.org/10.1016/j.ceramint.2019.01.029 741:measured the temperature of the atmosphere of 1908: 8: 1980:Vibrational spectroscopy of linear molecules 1371:Handbook of Vibrational Spectroscopy, Vol 1 1128: 1126: 1124: 1122: 1120: 985:transform infrared spectroscopy is used in 180:ultraviolet-visible ("UV-vis") spectroscopy 91:An example of an FTIR spectrometer with an 2271: 1975:Nuclear resonance vibrational spectroscopy 1915: 1901: 1893: 1871:The Grubb-Parsons-NPL cube interferometer 757:himself attempted to resolve the hydrogen 447:Measuring and processing the interferogram 2348:Inelastic electron tunneling spectroscopy 2028:Resonance-enhanced multiphoton ionization 1838: 1828: 1779: 1714: 1634:The Journal of Physical Chemistry Letters 1608: 1589:The Journal of Physical Chemistry Letters 1559: 1317:Journal of the Optical Society of America 997:FTIR is also used to investigate various 360:Learn how and when to remove this message 285:because water vapour has an intense pure 253:made from a single crystal of rock-salt ( 150:Learn how and when to remove this message 2116:Extended X-ray absorption fine structure 1484:Journal of Food Composition and Analysis 1136:Fourier Transform Infrared Spectrometry 1116: 38:Fourier-transform infrared spectroscopy 18:Fourier Transform Infrared Spectroscopy 1040:scanning near-field optical microscopy 1703:Review of Palaeobotany and Palynology 7: 2421: 1429:International Urology and Nephrology 1168: 1166: 1164: 1162: 1160: 342:adding citations to reliable sources 132:adding citations to reliable sources 32:Frustrated total internal reflection 1369:Griffiths, P.R.; Holmes, C (2002). 489:) the shortest wavelength would be 44:) is a technique used to obtain an 1050:FTIR as detector in chromatography 827:Far-IR spectrometers commonly use 25: 2333:Deep-level transient spectroscopy 2085:Saturated absorption spectroscopy 2420: 2409: 2408: 2338:Dual-polarization interferometry 318: 108: 80:originates from the fact that a 2353:Scanning tunneling spectroscopy 2328:Circular dichroism spectroscopy 2323:Acoustic resonance spectroscopy 1224:. July 27, 2009. Archived from 1104:Least-squares spectral analysis 329:needs additional citations for 119:needs additional citations for 2282:Fourier-transform spectroscopy 1970:Vibrational circular dichroism 1716:10.1016/j.revpalbo.2023.105041 1098:Fourier-transform spectroscopy 684:generation-recombination noise 504:For a maximum path difference 460:is used). This can trigger an 1: 2080:Cavity ring-down spectroscopy 1985:Thermal infrared spectroscopy 1057:gel permeation chromatography 993:Nano and biological materials 964:region at about 750 nm. 2214:Inelastic neutron scattering 1410:10.1016/0020-0891(69)90023-2 1173:Krishnan, Kannan M. (2021). 881:Attenuated total reflectance 876:Attenuated total reflectance 870:Attenuated total reflectance 93:attenuated total reflectance 2275:Data collection, processing 2151:Photoelectron/photoemission 1884:University FTIR lab example 1646:10.1021/acs.jpclett.8b01675 1423:Singh, Iqbal (2008-09-01). 753:at 0.1 cm resolution. 747:vibration-rotation spectrum 462:analog-to-digital converter 2475: 2360:Photoacoustic spectroscopy 2302:Time-resolved spectroscopy 1496:10.1016/j.jfca.2023.105276 1086:Discrete Fourier transform 956:Near-infrared spectroscopy 953: 893:discrete Fourier transform 873: 371: 29: 2404: 2386:Astronomical spectroscopy 2365:Photothermal spectroscopy 1441:10.1007/s11255-007-9327-2 1252:Agilent Technologies, Inc 929:With the advent of cheap 899:(FFT) algorithm is used. 615: 416:constructive interference 686:). For a spectrum with 388:Michelson interferometer 374:Michelson interferometer 310:Michelson interferometer 291:Michelson interferometer 229:capable of recording an 203:Michelson interferometer 2370:Pump–probe spectroscopy 2259:Ferromagnetic resonance 2051:Laser-induced breakdown 1002:with the help of FTIR. 772:entrance and exit slits 408:optical path difference 176:absorption spectroscopy 100:Conceptual introduction 2449:Scientific instruments 2066:Glow-discharge optical 2046:Raman optical activity 1960:Rotational–vibrational 1866:Infracord spectrometer 1544:10.1038/nprot.2014.110 1356:10.1364/JOSA.58.000433 1329:10.1364/JOSA.56.000896 1006:Microscopy and imaging 897:fast Fourier transform 849: 829:pyroelectric detectors 803: 473: 431: 383: 171: 96: 2459:Infrared spectroscopy 2287:Hyperspectral imaging 1888:University of Bristol 1744:Nature Communications 847: 813:silicon carbide (SiC) 801: 788:Infrared spectroscopy 765:in the spectrum of a 676:signal-to-noise ratio 508:adjacent wavelengths 471: 429: 381: 165: 90: 77:infrared spectroscopy 2039:Coherent anti-Stokes 1994:UV–Vis–NIR "Optical" 1294:10.1364/AO.35.002891 672:Fellgett's advantage 338:improve this article 271:diffraction gratings 190:work, for example.) 188:UV–vis spectrometers 128:improve this article 2343:Hadron spectroscopy 2133:Conversion electron 2094:X-ray and Gamma ray 2001:Ultraviolet–visible 1821:2018Mate...11..586. 1756:2013NatCo...4.2890A 1402:1969InfPh...9..185C 1375:John Wiley and Sons 1286:1996ApOpt..35.2891B 1228:on January 13, 2017 729:spectral resolution 287:rotational spectrum 243:molecular vibration 225:The first low-cost 2391:Force spectroscopy 2316:Measured phenomena 2307:Video spectroscopy 2011:Cold vapour atomic 1878:Infrared materials 1830:10.3390/ma11040586 1764:10.1038/ncomms3890 850: 804: 474: 432: 398:and directed to a 384: 247:dispersing element 178:techniques (FTIR, 172: 97: 75:Fourier-transform 2436: 2435: 2400: 2399: 2292:Spectrophotometry 2219:Neutron spin echo 2193:Beta spectroscopy 2106:Energy-dispersive 1640:(14): 4059–4065. 1601:10.1021/jz300150v 1280:(16): 2891–2896. 1184:978-0-19-883025-2 1150:978-0-471-19404-0 1092:Fourier transform 936:helium–neon laser 887:Fourier transform 745:by recording the 721:Fourier transform 670:The multiplex or 655:rapid calculation 628: 627: 370: 369: 362: 259:potassium bromide 231:infrared spectrum 227:spectrophotometer 215:Fourier transform 207:wave interference 160: 159: 152: 82:Fourier transform 16:(Redirected from 2466: 2454:Fourier analysis 2424: 2423: 2412: 2411: 2272: 2183:phenomenological 1932:Vibrational (IR) 1917: 1910: 1903: 1894: 1853: 1852: 1842: 1832: 1800: 1794: 1793: 1783: 1735: 1729: 1728: 1718: 1694: 1688: 1681: 1675: 1672: 1666: 1665: 1629: 1623: 1622: 1612: 1580: 1574: 1573: 1563: 1538:(8): 1771–1791. 1532:Nature Protocols 1522: 1516: 1515: 1475: 1469: 1468: 1420: 1414: 1413: 1390:Infrared Physics 1385: 1379: 1378: 1366: 1360: 1359: 1339: 1333: 1332: 1312: 1306: 1305: 1269: 1263: 1262: 1260: 1259: 1244: 1238: 1237: 1235: 1233: 1214: 1208: 1207: 1200:Clinical Science 1195: 1189: 1188: 1170: 1155: 1154: 1139:(2nd ed.). 1130: 974:chemical imaging 917:and marketed by 645: 641: 637: 545: 544: 529: 525: 521: 514: 507: 496: 492: 488: 486: 480: 365: 358: 354: 351: 345: 322: 314: 170:to the detector. 155: 148: 144: 141: 135: 112: 104: 95:(ATR) attachment 21: 2474: 2473: 2469: 2468: 2467: 2465: 2464: 2463: 2439: 2438: 2437: 2432: 2396: 2374: 2311: 2263: 2225: 2197: 2139: 2089: 1989: 1950:Resonance Raman 1926: 1921: 1862: 1857: 1856: 1802: 1801: 1797: 1737: 1736: 1732: 1696: 1695: 1691: 1682: 1678: 1673: 1669: 1631: 1630: 1626: 1582: 1581: 1577: 1524: 1523: 1519: 1477: 1476: 1472: 1422: 1421: 1417: 1387: 1386: 1382: 1368: 1367: 1363: 1344:J. Opt. Soc. Am 1341: 1340: 1336: 1314: 1313: 1309: 1271: 1270: 1266: 1257: 1255: 1246: 1245: 1241: 1231: 1229: 1216: 1215: 1211: 1197: 1196: 1192: 1185: 1172: 1171: 1158: 1151: 1141:Wiley-Blackwell 1132: 1131: 1118: 1113: 1082: 1073: 1065: 1052: 1036: 1008: 995: 982: 970:process control 958: 952: 940:James W. Brault 927: 910: 905: 895:is needed. The 889: 878: 872: 864: 842: 835: 825: 809: 796: 780: 762: 752: 717: 664: 643: 639: 635: 623: 619: 610: 603: 595: 591: 585: 581: 572: 565: 557: 551: 537: 533: 527: 523: 520: 516: 513: 509: 505: 494: 490: 484: 482: 478: 449: 376: 366: 355: 349: 346: 335: 323: 312: 255:sodium chloride 223: 156: 145: 139: 136: 125: 113: 102: 35: 28: 23: 22: 15: 12: 11: 5: 2472: 2470: 2462: 2461: 2456: 2451: 2441: 2440: 2434: 2433: 2431: 2430: 2418: 2405: 2402: 2401: 2398: 2397: 2395: 2394: 2388: 2382: 2380: 2376: 2375: 2373: 2372: 2367: 2362: 2357: 2356: 2355: 2345: 2340: 2335: 2330: 2325: 2319: 2317: 2313: 2312: 2310: 2309: 2304: 2299: 2294: 2289: 2284: 2278: 2276: 2269: 2265: 2264: 2262: 2261: 2256: 2251: 2246: 2245: 2244: 2233: 2231: 2227: 2226: 2224: 2223: 2222: 2221: 2211: 2205: 2203: 2199: 2198: 2196: 2195: 2190: 2185: 2180: 2175: 2174: 2173: 2168: 2166:Angle-resolved 2163: 2158: 2147: 2145: 2141: 2140: 2138: 2137: 2136: 2135: 2125: 2120: 2119: 2118: 2113: 2108: 2097: 2095: 2091: 2090: 2088: 2087: 2082: 2077: 2076: 2075: 2070: 2069: 2068: 2053: 2048: 2043: 2042: 2041: 2031: 2025: 2020: 2015: 2014: 2013: 2003: 1997: 1995: 1991: 1990: 1988: 1987: 1982: 1977: 1972: 1967: 1962: 1957: 1952: 1947: 1942: 1936: 1934: 1928: 1927: 1922: 1920: 1919: 1912: 1905: 1897: 1891: 1890: 1881: 1875: 1869: 1861: 1860:External links 1858: 1855: 1854: 1815:(4): 586–599. 1795: 1730: 1689: 1676: 1667: 1624: 1595:(7): 939–944. 1575: 1517: 1470: 1435:(3): 595–602. 1415: 1396:(4): 189–209. 1380: 1373:. Chichester: 1361: 1350:(3): 433–434. 1334: 1323:(7): 896–910. 1307: 1274:Applied Optics 1264: 1239: 1209: 1190: 1183: 1156: 1149: 1115: 1114: 1112: 1109: 1108: 1107: 1101: 1095: 1089: 1081: 1078: 1072: 1069: 1064: 1061: 1051: 1048: 1035: 1032: 1024:histopathology 1007: 1004: 994: 991: 981: 978: 954:Main article: 951: 948: 931:microcomputers 926: 923: 909: 906: 904: 903:Spectral range 901: 888: 885: 874:Main article: 871: 868: 862: 841: 838: 833: 824: 821: 808: 805: 795: 792: 779: 776: 760: 750: 749:of Venusian CO 716: 713: 708: 707: 703: 695: 682:is limited by 663: 660: 632: 631: 630: 629: 626: 625: 621: 617: 613: 612: 608: 605: 601: 597: 596: 593: 589: 586: 583: 579: 575: 574: 570: 567: 563: 559: 558: 555: 554:and d = (n+1)λ 552: 549: 535: 531: 518: 511: 497:). Because of 448: 445: 372:Main article: 368: 367: 326: 324: 317: 311: 308: 298:, such as the 279:Golay detector 263:caesium iodide 222: 219: 168:interferometer 158: 157: 116: 114: 107: 101: 98: 26: 24: 14: 13: 10: 9: 6: 4: 3: 2: 2471: 2460: 2457: 2455: 2452: 2450: 2447: 2446: 2444: 2429: 2428: 2419: 2417: 2416: 2407: 2406: 2403: 2392: 2389: 2387: 2384: 2383: 2381: 2377: 2371: 2368: 2366: 2363: 2361: 2358: 2354: 2351: 2350: 2349: 2346: 2344: 2341: 2339: 2336: 2334: 2331: 2329: 2326: 2324: 2321: 2320: 2318: 2314: 2308: 2305: 2303: 2300: 2298: 2295: 2293: 2290: 2288: 2285: 2283: 2280: 2279: 2277: 2273: 2270: 2266: 2260: 2257: 2255: 2252: 2250: 2247: 2243: 2240: 2239: 2238: 2235: 2234: 2232: 2228: 2220: 2217: 2216: 2215: 2212: 2210: 2207: 2206: 2204: 2200: 2194: 2191: 2189: 2186: 2184: 2181: 2179: 2176: 2172: 2169: 2167: 2164: 2162: 2159: 2157: 2154: 2153: 2152: 2149: 2148: 2146: 2142: 2134: 2131: 2130: 2129: 2126: 2124: 2121: 2117: 2114: 2112: 2109: 2107: 2104: 2103: 2102: 2099: 2098: 2096: 2092: 2086: 2083: 2081: 2078: 2074: 2071: 2067: 2064: 2063: 2062: 2059: 2058: 2057: 2054: 2052: 2049: 2047: 2044: 2040: 2037: 2036: 2035: 2032: 2029: 2026: 2024: 2023:Near-infrared 2021: 2019: 2016: 2012: 2009: 2008: 2007: 2004: 2002: 1999: 1998: 1996: 1992: 1986: 1983: 1981: 1978: 1976: 1973: 1971: 1968: 1966: 1963: 1961: 1958: 1956: 1953: 1951: 1948: 1946: 1943: 1941: 1938: 1937: 1935: 1933: 1929: 1925: 1918: 1913: 1911: 1906: 1904: 1899: 1898: 1895: 1889: 1885: 1882: 1879: 1876: 1874: 1870: 1867: 1864: 1863: 1859: 1850: 1846: 1841: 1836: 1831: 1826: 1822: 1818: 1814: 1810: 1806: 1799: 1796: 1791: 1787: 1782: 1777: 1773: 1769: 1765: 1761: 1757: 1753: 1749: 1745: 1741: 1734: 1731: 1726: 1722: 1717: 1712: 1708: 1704: 1700: 1693: 1690: 1687: 1680: 1677: 1671: 1668: 1663: 1659: 1655: 1651: 1647: 1643: 1639: 1635: 1628: 1625: 1620: 1616: 1611: 1606: 1602: 1598: 1594: 1590: 1586: 1579: 1576: 1571: 1567: 1562: 1557: 1553: 1549: 1545: 1541: 1537: 1533: 1529: 1521: 1518: 1513: 1509: 1505: 1501: 1497: 1493: 1489: 1485: 1481: 1474: 1471: 1466: 1462: 1458: 1454: 1450: 1446: 1442: 1438: 1434: 1430: 1426: 1419: 1416: 1411: 1407: 1403: 1399: 1395: 1391: 1384: 1381: 1376: 1372: 1365: 1362: 1357: 1353: 1349: 1345: 1338: 1335: 1330: 1326: 1322: 1318: 1311: 1308: 1303: 1299: 1295: 1291: 1287: 1283: 1279: 1275: 1268: 1265: 1253: 1249: 1243: 1240: 1227: 1223: 1219: 1213: 1210: 1205: 1201: 1194: 1191: 1186: 1180: 1176: 1169: 1167: 1165: 1163: 1161: 1157: 1152: 1146: 1142: 1138: 1137: 1129: 1127: 1125: 1123: 1121: 1117: 1110: 1105: 1102: 1099: 1096: 1093: 1090: 1087: 1084: 1083: 1079: 1077: 1070: 1068: 1062: 1060: 1058: 1049: 1047: 1045: 1041: 1033: 1031: 1029: 1025: 1021: 1017: 1013: 1005: 1003: 1000: 999:nanomaterials 992: 990: 988: 979: 977: 975: 971: 967: 963: 957: 950:Near-infrared 949: 947: 945: 941: 937: 932: 924: 922: 920: 919:Grubb Parsons 916: 907: 902: 900: 898: 894: 886: 884: 882: 877: 869: 867: 860: 856: 846: 840:Beam splitter 839: 837: 830: 822: 820: 818: 814: 806: 800: 793: 791: 789: 785: 777: 775: 773: 768: 764: 763:emission band 756: 748: 744: 740: 739:Janine Connes 735: 730: 726: 722: 714: 712: 704: 700: 699:monochromator 696: 693: 689: 685: 681: 680:photodetector 677: 673: 669: 668: 667: 661: 659: 656: 651: 649: 644:0.001 cm 614: 606: 599: 598: 587: 577: 576: 568: 561: 560: 553: 547: 546: 543: 542: 541: 540: 539: 502: 500: 491:1.266 μm 479:0.633 μm 470: 466: 463: 459: 453: 446: 444: 441: 436: 428: 424: 420: 417: 413: 409: 405: 401: 400:beam splitter 397: 394:radiator, is 393: 389: 380: 375: 364: 361: 353: 343: 339: 333: 332: 327:This section 325: 321: 316: 315: 309: 307: 305: 301: 297: 296:minicomputers 292: 288: 284: 280: 276: 272: 268: 264: 260: 256: 252: 248: 244: 240: 236: 232: 228: 220: 218: 216: 210: 208: 204: 199: 196: 195:monochromatic 191: 189: 185: 184:monochromatic 181: 177: 169: 164: 154: 151: 143: 133: 129: 123: 122: 117:This section 115: 111: 106: 105: 99: 94: 89: 85: 83: 79: 78: 71: 69: 65: 62: 58: 54: 50: 47: 43: 39: 33: 19: 2425: 2413: 2393:(a misnomer) 2379:Applications 2297:Time-stretch 2188:paramagnetic 2006:Fluorescence 1939: 1924:Spectroscopy 1812: 1808: 1798: 1747: 1743: 1733: 1706: 1702: 1692: 1679: 1670: 1637: 1633: 1627: 1592: 1588: 1578: 1535: 1531: 1520: 1487: 1483: 1473: 1432: 1428: 1418: 1393: 1389: 1383: 1370: 1364: 1347: 1343: 1337: 1320: 1316: 1310: 1277: 1273: 1267: 1256:. Retrieved 1254:. 2010-05-14 1251: 1242: 1230:. Retrieved 1226:the original 1221: 1212: 1203: 1199: 1193: 1174: 1135: 1074: 1066: 1053: 1037: 1010:An infrared 1009: 996: 983: 980:Applications 959: 928: 925:Mid-infrared 911: 908:Far-infrared 890: 879: 851: 826: 810: 783: 781: 718: 709: 691: 687: 665: 652: 633: 503: 495:7900 cm 475: 454: 450: 437: 433: 421: 407: 403: 385: 356: 347: 336:Please help 331:verification 328: 283:water vapour 235:Perkin-Elmer 224: 211: 200: 192: 174:The goal of 173: 146: 137: 126:Please help 121:verification 118: 74: 72: 64:spectrometer 41: 37: 36: 1965:Vibrational 944:calibration 734:corner-cube 648:apodization 640:0.5 cm 636:0.5 cm 412:retardation 404:retardation 70:at a time. 68:wavelengths 2443:Categories 2171:Two-photon 2073:absorption 1955:Rotational 1868:photograph 1709:: 105041. 1684:8037-8044. 1490:: 105276. 1258:2023-11-04 1206:(2). 1957. 1111:References 1012:microscope 807:IR sources 794:Components 778:Motivation 725:wavenumber 715:Resolution 662:Advantages 611:= (n+1)/d 522:will have 458:HeNe laser 396:collimated 392:black-body 239:wavenumber 61:dispersive 53:absorption 2249:Terahertz 2230:Radiowave 2128:Mössbauer 1886:from the 1809:Materials 1772:2041-1723 1725:0034-6667 1552:1750-2799 1512:257530876 1504:0889-1575 1449:1573-2584 1044:nano-FTIR 966:Overtones 823:Detectors 755:Michelson 620:− ν 573:=d/(n+1) 350:June 2022 275:bolometer 267:microwave 140:June 2022 73:The term 2415:Category 2144:Electron 2111:Emission 2061:emission 2018:Vibronic 1849:29641451 1790:24301518 1750:: 2890. 1662:49621115 1654:29957958 1619:22563521 1570:24992094 1457:18228157 1302:21085438 1232:March 5, 1080:See also 1030:grains. 946:errors. 767:hydrogen 499:aliasing 487: cm 233:was the 57:emission 49:spectrum 46:infrared 2427:Commons 2254:ESR/EPR 2202:Nucleon 2030:(REMPI) 1840:5951470 1817:Bibcode 1781:3863900 1752:Bibcode 1610:3341589 1561:4480339 1465:2249696 1398:Bibcode 1282:Bibcode 1222:Agilent 987:geology 962:visible 221:History 2268:Others 2056:Atomic 1847:  1837:  1788:  1778:  1770:  1723:  1660:  1652:  1617:  1607:  1568:  1558:  1550:  1510:  1502:  1463:  1455:  1447:  1300:  1181:  1147:  1028:pollen 1020:pixels 832:(LiTaO 817:Globar 727:. The 624:= 1/d 548:d = nλ 304:Varian 249:was a 2209:Alpha 2178:Auger 2156:X-ray 2123:Gamma 2101:X-ray 2034:Raman 1945:Raman 1940:FT-IR 1658:S2CID 1508:S2CID 1461:S2CID 1016:array 859:KRS-5 743:Venus 607:and ν 604:= n/d 592:= 1/λ 588:and ν 582:= 1/λ 569:and λ 566:= d/n 534:and ν 528:(n+1) 386:In a 300:PDP-8 251:prism 1845:PMID 1786:PMID 1768:ISSN 1721:ISSN 1650:PMID 1615:PMID 1566:PMID 1548:ISSN 1500:ISSN 1453:PMID 1445:ISSN 1298:PMID 1234:2013 1179:ISBN 1145:ISBN 972:and 653:For 526:and 515:and 42:FTIR 2237:NMR 1835:PMC 1825:doi 1776:PMC 1760:doi 1711:doi 1707:321 1642:doi 1605:PMC 1597:doi 1556:PMC 1540:doi 1492:doi 1488:119 1437:doi 1406:doi 1352:doi 1325:doi 1290:doi 915:NPL 857:or 855:CsI 784:why 485:800 440:KBr 406:or 340:by 130:by 55:or 51:of 2445:: 2242:2D 2161:UV 1843:. 1833:. 1823:. 1813:11 1811:. 1807:. 1784:. 1774:. 1766:. 1758:. 1746:. 1742:. 1719:. 1705:. 1701:. 1656:. 1648:. 1636:. 1613:. 1603:. 1591:. 1587:. 1564:. 1554:. 1546:. 1534:. 1530:. 1506:. 1498:. 1486:. 1482:. 1459:. 1451:. 1443:. 1433:40 1431:. 1427:. 1404:. 1392:. 1348:58 1346:. 1321:56 1319:. 1296:. 1288:. 1278:35 1276:. 1250:. 1220:. 1204:16 1202:. 1159:^ 1143:. 1119:^ 976:. 790:. 774:. 538:: 483:15 306:. 1916:e 1909:t 1902:v 1851:. 1827:: 1819:: 1792:. 1762:: 1754:: 1748:4 1727:. 1713:: 1664:. 1644:: 1638:9 1621:. 1599:: 1593:3 1572:. 1542:: 1536:9 1514:. 1494:: 1467:. 1439:: 1412:. 1408:: 1400:: 1394:9 1377:. 1358:. 1354:: 1331:. 1327:: 1304:. 1292:: 1284:: 1261:. 1236:. 1187:. 1153:. 863:2 834:3 761:α 759:H 751:2 692:m 688:m 622:1 618:2 616:ν 609:2 602:1 600:ν 594:2 590:2 584:1 580:1 578:ν 571:2 564:1 562:λ 556:2 550:1 536:2 532:1 524:n 519:2 517:λ 512:1 510:λ 506:d 493:( 481:( 363:) 357:( 352:) 348:( 334:. 153:) 147:( 142:) 138:( 124:. 40:( 34:. 20:)

Index

Fourier Transform Infrared Spectroscopy
Frustrated total internal reflection
infrared
spectrum
absorption
emission
dispersive
spectrometer
wavelengths
infrared spectroscopy
Fourier transform

attenuated total reflectance

verification
improve this article
adding citations to reliable sources
Learn how and when to remove this message

interferometer
absorption spectroscopy
ultraviolet-visible ("UV-vis") spectroscopy
monochromatic
UV–vis spectrometers
monochromatic
Michelson interferometer
wave interference
Fourier transform
spectrophotometer
infrared spectrum

Text is available under the Creative Commons Attribution-ShareAlike License. Additional terms may apply.